99
HANDPIECE FINE ADJUSTMENT SETTINGS
Handpiece
Item Number
Type
Normal Operating
Range
Str
okes Per Minute
Normal Air
Pressure
Range
psi (bar)
Recommended Initial Setting
Strokes
Per Minute
Air Pressure
psi (bar)
Magnum
004-940
800-3400
18-22 psi (1.2-1.5 bar)
2400
20 psi (1.4 bar)
901
004-901,
004-910
Standar
d
1400-4000
17-22 psi (1.2-1.5 bar)
2400
19 psi (1.3 bar)
Fine
1400-4000
12-15 psi (0.8-1.0 bar)
2400
13 psi (0.9 bar)
Monarch
004-921,
004-926
Standar
d
2000-5000
10-13 psi (0.7-0.9 bar)
3000
11 psi (0.8 bar)
Fine
2000-5000
6-8 psi (0.4-0.6 bar)
3000
7 psi (0.5 bar)
Maestro MX
004-909
600-3200
18-22 psi (1.2-1.5 bar)
2200
20 psi (1.4 bar)
Maestro EX
004-905
800-3600
17-22 psi (1.2-1.5 bar)
2200
19 psi (1.3 bar)
Maestro
004-947
2000-5000
10-13 psi (0.7-0.9 bar)
3000
11 psi (0.8 bar)
QC 720
004-720
400-3000
18-26 psi (1.2-1.8 bar)
1600
22 psi (1.5 bar)
QC 710
004-710
800-3000
20-24 psi (1.4-1.7 bar)
1800
21 psi (1.4 bar)
610 Hammer
004-610,
004-609
800-3000
20-24 psi (1.4-1.7 bar)
1800
21 psi (1.4 bar)
QC 801
004-801,
004-810
(obsolete)
1400-4000
20-24 psi (1.4-1.7 bar)
2400
22 psi (1.5 bar)
506 Large
004-506
(obsolete)
400-3000
18-26 psi (1.2-1.8 bar)
1600
22 psi (1.5 bar)
508 Standard
004-508
(obsolete)
800-3000
18-22 psi (1.2-1.5 bar)
2000
20 psi (1.4 bar)
Alter
natively
, the settings in the chart may be used for adjusting the selected handpiece; this method is not as pr
ecise.
Set the SPM dial and the air pr
essur
e contr
ol knob to the
Recommended Initial Setting
for the selected handpiece.