Troubleshooting 

■ Troubleshooting 

 

Tub fill doesn’t work even though the function is active and the 
light is flashing. 

Check all the sieves and valves for dirt. If the problem still ex-
ists, then the electronic has to be checked. Contact 

your GK rep-

resentative

.

 

 

The temperature regulation doesn’t work properly. 

The thermostatic mixer is adjusted at 3 bar water pressure at both 
supplies. If you have different water pressure it may be necessary 
to do a readjustment. If the water pressure is about right, possi-
bly the cartridge is defective and has to be changed. 
 

Very little water runs into the tub. 

If the hot and cold water supply pressure deviate too much from 
each other, then the mixing valve cannot work properly.

 

 
Disinfection doesn’t work and the button is flashing very fast 
(about 4x/sec.). 

The disinfectant container is empty. Replace it with a full disin-
fection container. 

 
Disinfection doesn’t work and the button is flashing very fast 
(about 4x/sec.) even though the container is full. 

The disinfection sensor is defective and has to be replaced.

 

 
All the functions are disabled except the “UP” and “DOWN” buttons. 

Probably the tub is still in the disinfection cycle. Finish the 
procedure properly. 
 

No shampoo or bath additive is dispensed even though the button is 
flashing when it is pressed. 

If the corresponding box is not empty, then there probably is air 
in the tubes. Press the corresponding button for a longer time. 

 

The hydromassage/ultrasound button doesn’t light up (with hydro-
massage/ultrasound being installed) even though there is more than 
enough water in the tub.  

The fill level sensor is defective and has to be changed. 

 

The tub fill does not stop automatically. 

The fill level sensor is defective and has to be changed.

 

 

All the functions are disabled except the “DOWN” button and the 
symbol “MISSING POWER SUPPLY” is lit. 

No power supply detected. Try to solve this problem as soon as 
possible since otherwise the emergency battery will be completely 
emptied.

 

Содержание GK Magic 1000

Страница 1: ......

Страница 2: ...and translation of both the whole document as well as of any parts thereof Congratulations on your decision to purchase a GK Bathtub You have acquired an Austrian quality prod uct which will considera...

Страница 3: ...er 8 Setting the temperature 9 Temperature display tub fill shower 9 Underwater light 9 Sound Vision 10 Ultrasound 11 Ultrasound cleaning FAQ 11 13 Hydromassage 14 Shampoo Bath additive 14 Disinfectio...

Страница 4: ...ce etc Never leave the patient unattended inside the bathtub Ensure that the equipment is thoroughly cleaned after every use When using disinfection protect your skin and eyes Heed all instructions on...

Страница 5: ...4 3 2 1 8 11 10 5 6 7 Components Components 1 Disinfection 2 Lid 3 Shower 4 Control panel 5 Mixer 6 Shampoo Bath additive dispenser 7 Fill in spout 8 Overflow 9 Hydromassage jets 10 Drain 11 Grab han...

Страница 6: ...5 Components Hydromassage jet Box for care and disinfection liquids USB port Flowmeter with con centration table...

Страница 7: ...Sensor Activation card 6 Button Down 7 Button Up 8 Button Tub fill 9 Button Shower 10 Button Bath additive 11 Button Disinfection 12 Button Shampoo 13 Button Hydromassage Ultrasound 14 Button Underwa...

Страница 8: ...oment and locked Only the sensor ACTIVATION CARD is illuminated with a green backlight The standby mode a prevents unauthorized use of the bathtub and b helps to save energy since due to the deactivat...

Страница 9: ...n TUB FILL the tub starts filling The button starts flashing and the symbol TUB FILL TEMPERATURE lights up The water now flows into the bathtub until the maximum filling level has been reached or the...

Страница 10: ...MPERATURE is illuminated additionally to the flashing button TUB FILL If the shower is active the symbol SHOWER TEMPERATURE is illuminated additionally to the flash ing button SHOWER If tub fill and t...

Страница 11: ...on PLAY PAUSE during playback it is paused and the button starts to shine continu ously When pressing it again the playback contin ues In case the pause exceeds 30 seconds the player resets itself to...

Страница 12: ...ination of the button stops and it can not be used until the next bathing cycle Note The ultrasound transducer is audible The closer any body parts are to the transducer plate the louder the side nois...

Страница 13: ...and spread out through the bathing water What effect does ultrasound have regarding body cleaning Because of the ultrasound millions of tiny gas bubbles are created in the water which collapse again v...

Страница 14: ...take unnecessary risks regarding the life of others May the ultrasound cleaning function be applied to people with pacemakers We have no information about any disruption of a pace maker s functionalit...

Страница 15: ...massage hose coupling is installed then also the jet containing the coupling will be activated immediately without delay upon activation of the whole function Shampoo Bath additive Both of these prod...

Страница 16: ...s infection button is flashing fast 4x sec To start the disinfection procedure take the red hand shower from its holder and press the disinfection but ton The button is flashing now Spray all internal...

Страница 17: ...Always check the water temperature with your bare hand also when using the shower before bringing the water in contact with the patient To add bath additive to the water if installed available press t...

Страница 18: ...y damage caused by unapproved unsuitable products will not be covered by the manufacturer s warranty Dark grey and transparent plastic parts The dark grey and transparent plastic parts may only be cle...

Страница 19: ...h From time to time the hydromassage jet covers should be removed and be cleaned together with the inlet and outlet aper tures separately To do so unscrew the jet covers in counter clockwise direction...

Страница 20: ...infection doesn t work and the button is flashing very fast about 4x sec even though the container is full The disinfection sensor is defective and has to be replaced All the functions are disabled ex...

Страница 21: ...or other dam ages Caution Should you notice any damage to the bathtub either during maintenance or at any other time then the bathtub may not be used any further from that point onward until the damag...

Страница 22: ...0 C Max relative humidity 80 not condensed at 20 C Atmospheric pressure range 500 hPa to 1060 hPa Transport and Storage Ambient temperature range 20 C to 60 C Relative humidity range 80 not condensed...

Страница 23: ...on class I Degree of protection IPX4 Protection type Type B Lifting weight 400 kg Lifting height 400 mm Maximum length 2240 mm Maximum width 930 mm Maximum height 1250 1650 mm Time of guarantee 2 year...

Страница 24: ...ents All product related information is not subject to change for technical production or other reasons Georg Kr mer Ges m b H Co KG A 4061 Pasching Neuhauserweg 14 Tel 43 0 7229 64450 Fax 43 0 7229 7...

Отзывы: