Generally output characteristic of any load cell has non-linear parts in the
beginning and in the end. That is why it is good to use sensor with bigger capacity than
scale capacity. Optimal is to use a load cell with 2 to 3 times higher capacity. That way
its guaranteed that these non-linear stages will not be part of measurement range.
4.Menu „Other”
In menu „Other” are placed parameters which rarely should be set or changed,
even from service engineers. These parameters:
Fact
- set all parameters and calibrations to factory default values
Enter this menu by pressing button
. Since here is only one parameter the only
buttons that work, are buttons
and
, respectively confirm or Exit.
4.1. Parameter „Fact”
With this option entire non-volatile memory of G5102 can be erased and all
parameters will be set to factory default values. It should not be used often – only when
the person setting the scale is in “dead end” and is wondering what to do with G5102.
Confirmation of erasing memory is on two levels - first button
should be
pressed to enter in this function. Then user will see blinking sign “
Attent
”, which means
„Attention! You are going to delete all settings!”. If user presses again button
, actual
erasing of whole non-volatile memory will begin. On display user will see sign
„-hold-“
,
which means that the user should wait process to finish and not turn off G5102 from
power grid until all is done.
Factory default settings for all important parameters are:
Number of metrological divisions
-
n = 3000
Weight
division
value
-
е
= 1
Decimal point position
-
dPP = 1
Zero
track
speed -
Ztrac = 0.5
e
/1s
Range of button
-
ZPerc = ±2% F.S
Range of button
-
tPerc = 50% F.S
Overload
-
oL = MAX + 8e
ADC amplifying
-
А
dc A = 1
Sampling rate for 1 measurement
-
t-me = 150ms
Bits for shifting right the result
-
Shr = 6 bit-a
Zero position
-
3000 divisions of ADC
Full range swing in divisions
-
20 000 divisions of ADC
Fine correction
-
0
е
in Full scale
Serial
interface
speed
-
baud = 9600bps, None
Scale serial number and firmware version of the scale, of course, are not deleted.