changes with every
“Arrow Down”
or
“Arrow Up”
press with pre-defined values.
Indication that user is in parameter value change mode are LEDs for “Tare” and “Zero”
position – they both are lit.
With these parameters sensitivity and measurement capabilities of
G5102 unit can be changed! Have this in mind when setting and
calibrating device!
Graphically process of setting parameters in menu
„Adc”
is showed on Fig. 6:
Fig. 6. Algorithm to change parameter values in menu „Adc”
3.Menu
„Cal”
In this menu real weight calibrating is carried out. Here user can set the scale to
measure with desired precision in selected measurement range (nominal capacity of the
scale) in order to comply with metrological and user requirements.
Weight calibration can be done with any known mass – what is put on scale
platform is entered during the process by G5102 keyboard. Calibration is carried in two
Scale indicator G5102, v.1.1 – USER MANUAL