TECHNICAL CHARACTERISTICS 

 

TYPOLOGY (MDD 93/42/EEC) 

Class IIa Medical Device 

MODEL 

SUPER VEGA BATTERIA 

UNI EN ISO 10079-1 

HIGH VACUUM / LOW FLOW 

POWER FEEDING 

230V~ / 50Hz or Internally powered equipment  

(PB Battery 12V 3Ah) or 12Vdc (1.9A) from car battery 

POWER CONSUMPTION 

50VA 

FUSE 

F 1 x 630mA 250V 

MAXIMUM SUCTION PRESSURE (without jar) 

-75kPa (-0.75 Bar) 

MINIMUM SUCTION PRESSURE (without jar) 

Less – 40kPa (-0.40 bar) 

MAXIMUM SUCTION FLOW (without jar) 

16 l/min 

WEIGHT 

4.9 Kg 

SIZE 

350 x 210 x 180mm 

BATTERY HOLDING TIME 

45 ÷ 50 minutes 

BATTERY TIME CHARGE 

120 ÷ 150 minutes 

WORKING CONDITION 

Room temperature:                       5 ÷ 35°C 
Room humidity percentage:         30 ÷ 75% RH 
Altitude:                                           0 ÷ 2000m s.l.m. 

CONSERVATION CONDITION AND TRASPORT 

Room temperature:                       -40÷ 70°C 
Room humidity percentage:          10 ÷ 100% RH 

 

SYMBOLS 

 

 

Class II isolation equipment 

 

CE marking in conformity with EC directive 93/42/EEC 

Manuafcturer: CA.MI. di Attolini Mario & C. snc 

Via Ugo La Malfa nr.31 – 43010 Pilastro (PR) Italy 

 

Warning, consult the instruction manual 

 

To Preserve in place coolness and dry land 

 

Conservation temperature: -40 ÷ 70°C 

 

Type B equipment 

 

Fuse 

Alternate Current 

 

Battery 

 

Direct Current 

Hz 

Mains Frequency 

ON 

OFF 

 

ACCESSORIES SUPPLIED 

 

DESCRIPTION 

COMPLETE ASPIRATION JAR 1000cc 

CONICAL FITTING 

TUBES SET 6 mm x 10 mm 

ASPIRATION PROBE CH20 

ANTIBACTERIAL FILTER 

 
The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit. When the filter is wet, it’s not possible to use 
the unit therefore the filter should be changed immediately. In case of possible contamination or discolouration, change the filter immediately.  
Don’t use the suction unit without the protection filter fitted. If the suction unit is used in an emergency or in a patient where the risk of contamination is not 
know the filter must be changed after each use. Available under request with different versions with complete jar 2000cc.  

The suction probe must be changed at the end of each application. 
 
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standards 
on material biocompatibility. 

Содержание SUPER VEGA BATTERIA

Страница 1: ...aitaly com INTERNATIONAL Tel 39 02 953854209 221 225 Fax 39 02 95380056 e mail export gimaitaly com www gimaitaly com ASPIRATORE SUPER VEGA BATTERIA ASPIRATOR SUPER VEGA BATTERIA ASPIRATEUR SUPER VEGA...

Страница 2: ...il dispositivo medico sempre con il filtro antibatterico Non immergere mai l apparecchio in acqua Posizionare l apparecchio su superfici piane e stabili Posizionare l apparecchio in modo da evitare d...

Страница 3: ...35 C Percentuale umidit ambiente 30 75 RH Altitudine 0 2000m s l m CONDIZIONI DI CONSERVAZIONE E TRASPORTO Temperatura ambiente 40 70 C Percentuale umidit ambiente 10 100 RH SIMBOLOGIA ADOTTATA SULL A...

Страница 4: ...ospedale o ambienti alimentati da batterie Surge IEC EN 61000 4 5 1kV modo differenziale L alimentazione dovrebbe essere quella tipica di un ambiente commerciale o ospedale o ambienti alimentati da b...

Страница 5: ...segni di contaminazione esterne e o interne render l apparecchio al cliente con chiaramente specificato APPARECCHIO NON RIPARATO allegando una lettera di spiegazioni sui difetti riscontrati GIMA giud...

Страница 6: ...8 ore luce gialla e poi verde 2 Luce rossa accesa Batteria scarica E possibile far funzionare il dispositivo alimentandolo tramite cavo elettrico inserito a rete luce gialla poi verde 3 Mancata aspir...

Страница 7: ...3 di acqua per facilitare le operazioni di pulizia e velocizzare la depressione durante il funzionamento quindi riavvitare il coperchio al vaso Estrarre gli accessori e procedere alle operazioni di p...

Страница 8: ...ly with the bacteriological filter Never immerge the appliance into water Position the appliance on flat stable surfaces Position the device in a way that the air inlets on the back aren t obstructed...

Страница 9: ...he instruction manual To Preserve in place coolness and dry land Conservation temperature 40 70 C Type B equipment Fuse Alternate Current Battery Direct Current Hz Mains Frequency I ON O OFF ACCESSORI...

Страница 10: ...25 cycle 5 UT for 5 sec Mains power quality should be that of a typical commercial environment or hospital If the user of the surgical aspirator SUPER VEGA BATTERIA request that the appliance operates...

Страница 11: ...ght switches over to the fixed green light Ensure that the green light comes on and that after a few seconds it switches over to the yellow light meaning that the charged battery is operating A protec...

Страница 12: ...to full position is of around 45 50 minutes with operation in continuous In the case in which instead the device is furnished with internal battery discharge the time of recharge of the internal batte...

Страница 13: ...les accessoires et les composants originaux Ne jamais plonger l appareil dans l eau Positionner l appareil sur des surfaces planes et stables Positionner l appareil de fa on viter d en occlure les pr...

Страница 14: ...continu I Allum O Eteint ELIMINATION DES BATTERIES USAG S Directrive 2006 66 CE Applicable dans les pays de l Union Europ enne et aux autres pays europ ens disposant de syst me de collecte selective C...

Страница 15: ...ff rentiel L alimentation devrait tre celle typique d un environnement commercial ou hospitalier or alimentation des environnements vous des batteries Trous de tension br ves interruptions et variatio...

Страница 16: ...LLAGES ET DES OPERATEURS QUI LES UTILISENT GIMA S p A COMPTE SUR LE RESPECT DES CET NORMES POUR POUVOIR GARATIR L HYGIENE ET LA SANTE A TOTES LES PERSONNES QUI OPERENT POUR OBTENIR LA QUALITE ET LE BI...

Страница 17: ...rg e On peut faire fonctionner l appareil en l alimentant avec le cordon lectrique branch au secteur lumi re jaune puis verte 3 L appareil n aspire pas Couvercle du bocal mal viss D visser et revisser...

Страница 18: ...ur sur la position 0 et d brancher la fiche de la prise d alimentation Sortir les accessoires et proc der aux op rations de nettoyage comme indiqu dans le chapitre nettoyage ATTENTION Le c te du filtr...

Страница 19: ...H nden anfassen Auf jeden Fall vermeiden dass es mit Fl ssigkeiten in Ber hrung kommt Unbedingt vermeiden dass Kinder und oder Behinderte das Ger t ohne die erforderliche berwachung benutzen Das Ger...

Страница 20: ...solierstoffklasse II CE Kennzeichnung gem Richtlinie 93 42 EWG Mit CA MI di Attolini Mario C snc Via Ugo La Malfa nr 31 43010 Pilastro PR Italia Achtung Im Handbuch nachlesen K hl und trocken lagern L...

Страница 21: ...000 4 3 3V m 80MHz to 2 5 GHz f r Ger te die nicht der Lebenserhaltung dienen Anmerkung UT ist der Wert der Einspeisungsspannung VORSCHRIFTEN F R ZUR CKSCHICKEN UND REPARIEREN In bereinstimmung mit de...

Страница 22: ...unter 40 kPa 0 40 bar abf llt Sicherstellen dass man keine st renden Ger usche h rt die auf einen Schaden hinweisen k nnten Es ist der Betrieb mit in die Steckdose gestecktem Netzstecker m glich die g...

Страница 23: ...Ger ts entladen ist nach die Benutzung muss man zum Laden der Batterie das Ger tekabel f r circa 120 150 Minuten mit ausgeschaltetem Hauptschalter des Ger ts an das Stromnetz von 230 V 50Hz anschlie e...

Страница 24: ...les El aparato solo se puede utilizar con filtros bacteriologicos Nunca sumergir el aparato en agua Posicionar el aparato sobre superficies planas y estables Posicionar el aparato de modo tal de evita...

Страница 25: ...rabilidad bater a 45 50 minutos Tempo de recharga bater a 120 150 minutos Condiciones de funcionamiento Temperatura ambiente 5 35 C Porcentaje de humedad ambiente 30 75 RH Altitud 0 2000m s l m Condic...

Страница 26: ...un hospital o ambientes a usted de las bater as Surge IEC EN 61000 4 5 1kV modo differecial La alimentac on deber a sera quella tipica de un ambiente comercial o de un hospital o ambientes a usted de...

Страница 27: ...ERSONAS QUE OPERAN PARA OBTENER CALIDAD Y BENESTAR GIMA otorga a los equipos Super Vega Batteria una garant a de 24 meses a partir de la fecha de compre Para hacer v lida la garant a el cliente deber...

Страница 28: ...oras luz amarilla y despu s verde 2 Luz roja encendida Bater a agotada Es posible hacer funcionar el aparato suministr ndole a trav s de un cable el ctrico colocado en red luz amarilla luego verde 3 N...

Страница 29: ...n l quidos aspirados Si el aparato tiene la bater a interna agotada despu s la utilizaci n para poderla cargar es necesario conectar el cable del aparato con la red el ctrica 230V 50Hz por aprox 120 1...

Страница 30: ...CTION Connessione apparecchio Suction pump Inlet Connessione Vaso Jar Air Tube Montaggio Filtro Filter Assembling Mod TOBI HOSPITAL TOBI HOSPI PLUS HOSPI PLUS Cod 28237 TOBI CLINIC SUPER TOBI CARRELLA...

Отзывы: