Warning
1) Don’t near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for
magnetic resonance imaging, where the intensity of EM DISTURBANCES is high.
2) Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
3) Use of accessories, transducers and cables other than those specified or provided by the manufac-
turer of this equipment could result in increased electromagnetic emissions or decreased electromag-
netic immunity of this equipment and result in improper operation.
4) Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the device including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could
result.
5) Active medical devices are subject to special EMC precautions and they must be installed and used
in accordance with these guidelines.
Note:
• When the device is disturbed, the data measured may fluctuate, please measure repeatedly or in
another environment to ensure its accuracy.
Disposal: The product must not be disposed of along with other domestic waste. The users must
dispose of this equipment by bringing it to a specific recycling point for electric and electronic
equipment.