43
ENGLISH
9 SPECIFICATION
SENSOR OFF
The sensor is off (probe-off)
Manufacturer
Alternating current indicator light
Temperature limit
IP21
Covering Protection rate
Atmospheric pressure limit
Date of manufacture
Fragile, handle with care
Humidity limit
Recovery
This side up
Lot number
Keep in a cool, dry place
Authorized representative in the
European community
P/N
Material code
WEEE disposal
Medical Device complies
with Directive 93/42/EEC
Product code
Note: Your device may not contain all the following symbols.
SpO2 [see note 1]
Display range
0% ~ 100%
Measured range
0% ~ 100%
Accuracy [see note 2]
70% ~ 100%: ±2%;
0% ~ 69%: unspecified.
Resolution
1%
PR
Display range
30 bpm ~ 250 bpm
Measured range
30 bpm ~ 250 bpm
Accuracy [see note 3]
±2 bpm or ±2%, whichever is greater.
Resolution
1 bpm
PI
Display range
0% ~ 20%
Measured range
0% ~ 20%
Accuracy
1% ~ 20%: ±1%
0% ~ 0,9%: ±0,2%
Resolution
0,1%
Accuracy under low perfusion
[see note 4]
Low perfusion 0.4%:
SpO
2
: ±4%;
PR: ±2 bpm or ±2%, whichever is greater
Light interference
Under normal and ambient light conditions, the SpO
2
deviation ≤ 1%
Pulse intensity
Continuous bar graph display, the higher display indicates the strong-
er pulse.
Upper and lower limit of alarm values
SpO
2
0% ~ 100%
PR
0 bpm ~ 254 bpm