19
Performing a Quality Control Test
The quality control test confirms that the test strips and meter are working
together properly, and that you are performing the test correctly. It is important to
perform this test:
Before you first use your meter.
Before using a new box of test strips.
When you suspect that the meter or test strips are not working properly.
When you suspect that your test results are inaccurate, or if they are
inconsistent with how you feel.
When you suspect your meter is damaged.
After cleaning your meter.
At least once a week.
1. Insert a test strip into the strip port, contact bars end first and facing up, to turn
on the meter and display all the display segments. If the audio option is on,
the meter will beep, signaling the meter is turned on.
2. Check the display to confirm that all the display segments turn on. Next, a
dash will move across the display. See illustrations below.
3. The meter is ready for testing when the blinking blood drop and strip symbol
appear. The display will show the date time and the strip icon with the blood
sample icon blinking to indicate that the test strip is inserted correctly. You can
then add a drop of control solution.
Note
:
If the test strip has been inserted incorrectly, the meter will not turn on.
NOT READY NOT READY NOT READY
READY TO TEST