MB122-MB132-MB160 

Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones 

MA036IGBEa  

 

41

 

6.5 REPAIRS 

High frequency cables and electrode holder handle cannot be repaired. Always substitute a damaged 
part with a new one.  

6.5.1

 

F

USE 

S

UBSTITUTION

 

Before substituting the fuse, disconnect the unit from the mains system. 

Only use fuse of the kind 5x20; they must have those characteristics: 3.15AT (slow) (230Vac mains 
voltage), 6.3A (115Vac mains voltage), proceed as follows: 
1-2 Extract, with a small screwdriver, the drawer under the mains socket 
3 Extract the damaged fuse and insert two new ones 
4 Reinsert the drawer

 

 

6.6 CHECKING OF THE EQUIPMENT BEFORE EACH USE 

Each time the use of the electrosurgical equipment is planned a check of the most important safety 
aspects has to be implemented considering at least the following: 

 

Check the integrity of cords, connections, wires breakage, etc. 

 

Assure that all the electrical equipment is properly grounded 

 

Assure that all the accessories that should be used are available and sterilised. 

 

Check, by disconnecting the reference electrode cable, the functioning of the OC light and sound 
alarm warning. 

 

Check, by activating the CUT and COAG power switch, the functioning of the emission lights and 
sounds warnings. 

6.7 FUNCTION AND SAFETY CHECK AND TEST 

At least once a year, the biomedical engineering department or other qualified personnel should do the 
following check and test: 

 

Check of the connectors and mains supply cord conditions; 

 

Visual check of the mechanical protections; 

 

Check of the protections against the danger due to liquid's pouring, dripping, moisture, liquid's 
penetration, cleanliness, sterilization and disinfection. 

 

Check of the Equipment's Datas on the Label 

 

Check of the availability of the Instruction's Manual 

 

Check the functioning of the H.F. output controls 

 

Check the uniformity of the resistance through the surface of the patient plate. 

 

Test the earth conductivity resistance. 

 

Test the earth leakage current. 

 

Test H.F. leakage current. 

 

Control of the neuromuscular stimulation. 

 

Control of the accuracy of the output power. 

Содержание DIATERMO MB122

Страница 1: ...MANUALE D ISTRUZIONI INSTRUCTION S MANUAL MANUAL DE INSTRUCCIONES DIATERMO MB122 MB132 MB160 M30540 MA036IGBEa...

Страница 2: ...rsonnel at all times All the safety instructions and advice notes are to be observed Be sure that these operating instructions is furnished together the equipment when this is transferred to other ope...

Страница 3: ...D COAG 18 4 4 4 Corrente per Coagulazione Profonda SOFT COAG 18 4 4 5 Corrente di Coagulazione Bipolare BIPOLAR 18 4 5 SEGNALAZIONE DI TEMPO DI EROGAZIONE ECCESSIVO OVT 19 4 6 SEGNALAZIONE DI ECCESSIV...

Страница 4: ...on Current SOFT COAG 35 4 4 5 Bipolar Coagulating Current BIPOLAR 35 4 5 SIGNALING OF EXCESSIVE TIME OF DELIVERY OVT 36 4 6 SIGNALING OF EXCESSIVE IMPEDANCE IN THE CIRCUIT OF NEUTRAL ELECTRODE OC 36 4...

Страница 5: ...de la Coagulacion BIPOLAR 53 4 5 EL SE ALAR DE LA POCA EXCESIVA DE LA SALIDA OVT 54 4 6 EL SE ALAR DE LA IMPEDANCIA EXCESIVA EN EL CIRCUITO NEUTRAL DEL ELECTRODO OC 54 4 7 AJUSTE DEL NIVEL AC STICO DE...

Страница 6: ...room It has foreseen its use in the monopolar cut cut coagulated or coagulation mode or in bipolar coagulation mode The equipment is conceived for being used in the following sectors DIATERMO descrizi...

Страница 7: ...00500 L10 Bent ball electrode 3mm L10 5pcs 5 cm 30529 00500 L10 L Bent ball electrode 3mm L10 L 5pcs 10 cm 30505 00500 L5 Bent hook electrode L5 5pcs 5 cm 30525 00500 L5 L Bent hook electrode L5 L 5pc...

Страница 8: ...effects can be precisely controlled it can be used safely without damage to the bone but since good coagulation while cutting is one of principal benefits of using electrosurgery a current with a cert...

Страница 9: ...ectioning of the biological tissue due to the passage of HF current concentrated on the tips of the bipolar forceps When the HF current is applied to the tissue between the two tips of the forceps an...

Страница 10: ...le or metallic support The use of antistatic sheets is advised Avoid the skin to skin contact for example between arm and body of the patient Insert an interface material like dry surgical gauze Moreo...

Страница 11: ...e lack of efficient connection to earth of the installation Operation without a protective earth connection is forbidden Before connect the equipment ascertain that the required voltage showed on the...

Страница 12: ...aporate before putting in operation the unit Don t exceed the temperature environment or the allowed moisture Environmental conditions Temperature 10 40 C Relative moisture 30 75 Pressure 70 106k Pa W...

Страница 13: ...done pressing the correspondent pushbutton on the unit press the blue foot switch to pre set and deliver the coagulating current the choice between FORCED COAG SOFT COAG or BIPOLAR must be done pressi...

Страница 14: ...work 4 1 1 MANUFACTURER S IDENTIFICATION DATA DIATERMO MB122 MB160 and MB132 HF electrosurgical unit are designed manufactured and tested by the LED SpA in its own laboratories in Aprilia LT Italy 4 1...

Страница 15: ...ion keyboard 5 Alarm warning for current delivery over time 6 OC open circuit alarm 7 Handle connection 8 Return plate connection MB132 1 Mains switch 2 Pulse time selector 3 Pulse repetition selector...

Страница 16: ...t when a deep coagulation together the cut is desired This current is made by sine current e suited for the cut associated to low voltage current suited for coagulation soft coag With this blending a...

Страница 17: ...indications 1 Switch on the unit through the mains switch while the CUT pushbutton is maintained pressed 2 When the unit has finished to check internal parameters on the display CUT appears the messa...

Страница 18: ...push the CONT push button to do it 4 10 2 REPETITION OF TIMED DELIVERY The timed high frequency current emission can be automatically repeated from 2 to 9 times each time one foot switch or one hand s...

Страница 19: ...ided with line fuses WARNING before switch on the unit operator has to verify that requested mains voltage corresponds to the voltage available from the electrical net 4 11 2 PEDAL CONNECTOR On the le...

Страница 20: ...0 1050 1050 15 Maximum output voltage BLEND Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage FORCED Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage SOFT Vpp on5 2k9 540 540 540 15 Maximum out...

Страница 21: ...of the supplier of each item When original reusable accessories supplied by LED S p A are applied the cleaning by using soft cleanser and sterilization through steam sterilisation at 121 C is recommen...

Страница 22: ...s properly grounded Assure that all the accessories that should be used are available and sterilised Check by disconnecting the reference electrode cable the functioning of the OC light and sound alar...

Страница 23: ...edance load 100 2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nomina...

Страница 24: ...2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nominal value Diagrams...

Страница 25: ...end of the life the present product mustn t be eliminated as urban refusal but it must be eliminated in a separated collection If the product is eliminated in unsuitable way it is possible that some p...

Отзывы: