background image

GIMA SpA

 

    DIATERMO 

MB122-MB132-MB160 

36  

MA036IGBEa  

Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones

 

4.5 SIGNALING OF EXCESSIVE TIME OF DELIVERY (OVT) 

If the operator exceeds the maximum time of disbursement, recommended by the international norms, 
that is 10 seconds, the equipment produces a signal of warning consisting of bright intermittent signal 
OVT. If despite the signal of warning the operator insists in the continuous delivering, after a time 
depending from the type of current, and from the level of the same one, the signal of warning is 
transformed in impediment to the delivering of current that is signalled through the signal OVT 
constantly illuminated. The interdiction lasting of the current delivery depends from the previous 
conditions of delivery. 

  

MB122 – MB160 

MB132 

 

 

4.6 SIGNALING OF EXCESSIVE IMPEDANCE IN THE CIRCUIT OF 
NEUTRAL ELECTRODE (OC) 

For the meaning of this warning signal please refer to the previous description of the neutral electrode circuit. 

4.7 ADJUSTMENT OF THE ACOUSTIC SIGNAL LEVEL 

To modify the emission acoustic signal it is necessary to follow those indications: 
1.

 

Switch on the unit through the mains switch while the CUT pushbutton is maintained pressed 

2.

 

When the unit has finished to check internal parameters, on the display CUT appears the message SOU., 
while on the COAG one, the value of the preset level. Now, the CUT pushbutton can be released. 

3.

 

Through the COAG knob it is possible varying the emission acoustic level. During the variation 
the sound emitted by the unit corresponds to the preset level.  

4.   Press the CUT pushbutton to confirm the level.    

 

 

 

Level 

Sound emission until 1m 

distance from the frontal panel

1 55 

dBA 

2 60 

dBA 

3 65 

dBA 

4 70 

dBA 

5 75 

dBA 

4.8 AUTOMATIC CONTROL OF THE INTERNAL PARAMETERS 

The unit has an automatic control system of some of the internal parameters. When switched on, the control 
is indicated on the display through the message SEL FCh. If there are not errors, the message PAS Sed 
appears; if there are errors, Err 001 appears. See Guide to the Problems’ Solution for further information. 

4.9 CONNECTORS  

Connector for return plate  
This is the point of connection of the return plate or of the bipolar optional adapter (

REF

 00498.04) for 

BIPOLAR function.

  

 

Connector for handle  
This is the point of connection of electrode handle. In the case of use of handle without finger switch it 
shall be connected to the black buckle.

  

 

Содержание DIATERMO MB122

Страница 1: ...MANUALE D ISTRUZIONI INSTRUCTION S MANUAL MANUAL DE INSTRUCCIONES DIATERMO MB122 MB132 MB160 M30540 MA036IGBEa...

Страница 2: ...rsonnel at all times All the safety instructions and advice notes are to be observed Be sure that these operating instructions is furnished together the equipment when this is transferred to other ope...

Страница 3: ...D COAG 18 4 4 4 Corrente per Coagulazione Profonda SOFT COAG 18 4 4 5 Corrente di Coagulazione Bipolare BIPOLAR 18 4 5 SEGNALAZIONE DI TEMPO DI EROGAZIONE ECCESSIVO OVT 19 4 6 SEGNALAZIONE DI ECCESSIV...

Страница 4: ...on Current SOFT COAG 35 4 4 5 Bipolar Coagulating Current BIPOLAR 35 4 5 SIGNALING OF EXCESSIVE TIME OF DELIVERY OVT 36 4 6 SIGNALING OF EXCESSIVE IMPEDANCE IN THE CIRCUIT OF NEUTRAL ELECTRODE OC 36 4...

Страница 5: ...de la Coagulacion BIPOLAR 53 4 5 EL SE ALAR DE LA POCA EXCESIVA DE LA SALIDA OVT 54 4 6 EL SE ALAR DE LA IMPEDANCIA EXCESIVA EN EL CIRCUITO NEUTRAL DEL ELECTRODO OC 54 4 7 AJUSTE DEL NIVEL AC STICO DE...

Страница 6: ...room It has foreseen its use in the monopolar cut cut coagulated or coagulation mode or in bipolar coagulation mode The equipment is conceived for being used in the following sectors DIATERMO descrizi...

Страница 7: ...00500 L10 Bent ball electrode 3mm L10 5pcs 5 cm 30529 00500 L10 L Bent ball electrode 3mm L10 L 5pcs 10 cm 30505 00500 L5 Bent hook electrode L5 5pcs 5 cm 30525 00500 L5 L Bent hook electrode L5 L 5pc...

Страница 8: ...effects can be precisely controlled it can be used safely without damage to the bone but since good coagulation while cutting is one of principal benefits of using electrosurgery a current with a cert...

Страница 9: ...ectioning of the biological tissue due to the passage of HF current concentrated on the tips of the bipolar forceps When the HF current is applied to the tissue between the two tips of the forceps an...

Страница 10: ...le or metallic support The use of antistatic sheets is advised Avoid the skin to skin contact for example between arm and body of the patient Insert an interface material like dry surgical gauze Moreo...

Страница 11: ...e lack of efficient connection to earth of the installation Operation without a protective earth connection is forbidden Before connect the equipment ascertain that the required voltage showed on the...

Страница 12: ...aporate before putting in operation the unit Don t exceed the temperature environment or the allowed moisture Environmental conditions Temperature 10 40 C Relative moisture 30 75 Pressure 70 106k Pa W...

Страница 13: ...done pressing the correspondent pushbutton on the unit press the blue foot switch to pre set and deliver the coagulating current the choice between FORCED COAG SOFT COAG or BIPOLAR must be done pressi...

Страница 14: ...work 4 1 1 MANUFACTURER S IDENTIFICATION DATA DIATERMO MB122 MB160 and MB132 HF electrosurgical unit are designed manufactured and tested by the LED SpA in its own laboratories in Aprilia LT Italy 4 1...

Страница 15: ...ion keyboard 5 Alarm warning for current delivery over time 6 OC open circuit alarm 7 Handle connection 8 Return plate connection MB132 1 Mains switch 2 Pulse time selector 3 Pulse repetition selector...

Страница 16: ...t when a deep coagulation together the cut is desired This current is made by sine current e suited for the cut associated to low voltage current suited for coagulation soft coag With this blending a...

Страница 17: ...indications 1 Switch on the unit through the mains switch while the CUT pushbutton is maintained pressed 2 When the unit has finished to check internal parameters on the display CUT appears the messa...

Страница 18: ...push the CONT push button to do it 4 10 2 REPETITION OF TIMED DELIVERY The timed high frequency current emission can be automatically repeated from 2 to 9 times each time one foot switch or one hand s...

Страница 19: ...ided with line fuses WARNING before switch on the unit operator has to verify that requested mains voltage corresponds to the voltage available from the electrical net 4 11 2 PEDAL CONNECTOR On the le...

Страница 20: ...0 1050 1050 15 Maximum output voltage BLEND Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage FORCED Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage SOFT Vpp on5 2k9 540 540 540 15 Maximum out...

Страница 21: ...of the supplier of each item When original reusable accessories supplied by LED S p A are applied the cleaning by using soft cleanser and sterilization through steam sterilisation at 121 C is recommen...

Страница 22: ...s properly grounded Assure that all the accessories that should be used are available and sterilised Check by disconnecting the reference electrode cable the functioning of the OC light and sound alar...

Страница 23: ...edance load 100 2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nomina...

Страница 24: ...2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nominal value Diagrams...

Страница 25: ...end of the life the present product mustn t be eliminated as urban refusal but it must be eliminated in a separated collection If the product is eliminated in unsuitable way it is possible that some p...

Отзывы: