3. INSTALLATION
•
Inspect the unit for damages during transport. The claims for possible damages will be accepted
only in case they are immediately communicated to the carrier; the damages that are found must
be written down and presented to LED SpA or to your own retailer. If the unit is returned to the
LED SpA or to your own retailer, it is necessary to use the original equipment’s package or
another equivalent one, to guarantee the safety during the transport.
•
Unpack the equipment and carefully study the documentation and operating instruction supplied.
Mains voltage, indicated above the inlet, must agree with the local mains voltage (mains voltage
frequency: 50-60 Hz). The correct voltage setting is selected by turning the voltage selector when
available. Insert the correct fuses in the module referring to the value written on the label.
•
Connect mains cable to a mains outlet having good hearth connection
OPERATION OF THE EQUIPMENT WITHOUT EARTH CONNECTION IS
FORBIDDEN
•
The unit must be installed on a level surface, with dimension, at least, correspondent to those of
the base of the unit itself. Around the unit must be left a space of 25cm, at least.
•
Connect the mains cable to the mains socket on the rear panel of the unit.
•
Connect the equipotential binding post located at the left of the unit’s back panel to equipotential
socket of the plant.
•
Connect the single foot switch or the double foot switch (optional) to the connector on the rear
panel.
•
Connect handle, in the case of use of handle without finger switch it shall be connected on the
black buckle.
•
In case of use of bipolar forceps (see BIPOLAR operation paragraph 4.4.5) it is necessary to use
the special optional adapter (
REF
00498.04).
•
Let unit work in dry environment only. Any verified condensate must be let evaporate before
putting in operation the unit. Don't exceed the temperature environment or the allowed moisture.
Environmental conditions: Temperature: 10/40°C - Relative moisture: 30/75% - Pressure:
70/106k
Pa
•
When the unit is switched on, through the on/off switch on the frontal panel, after having checked
the internal parameters, it will work with the function and the power level utilized during the last
switching (when the unit is switched for the first time the level will be 00).
•
Before using the unit, it is necessary connect the cable to the patient plate. Single plate electrodes
and split plate electrodes can be. If the value of the impedance is acceptable, the OC indicator light
will stop flashing and the alarm to sound.
•
Having:
Holder-handle with two pushbuttons without foot switch
: press the yellow pushbutton on the holder-
handle to deliver the cutting current (the choice between CUT or BLEND must be done pressing the
correspondent pushbutton on the unit); or the blue pushbutton on the holder handle to deliver
coagulating current (the choice between FORCED COAG, SOFT COAG or BIPOLAR must be done
pressing the correspondent pushbutton on the unit).
Holder handle with two pushbuttons and a single foot switch
: choose the cutting current CUT or
BLEND and the coagulation current FORCED COAG, SOFT COAG or BIPOLAR. Preset through the
yellow pushbutton on the holder handle, the function for the cut that appears on the unit or, through the
blue pushbutton on the holder handle, the function for the coagulation that appears on the unit. The
current delivery takes place through the foot switch.
Holder handle with two pushbuttons and double foot switch
: press the yellow foot switch or the
yellow pushbutton of the holder handle to pre-set and deliver the cutting current (the choice between
CUT or BLEND must be done pressing the correspondent pushbutton on the unit) or the blue foot
switch or the blue pushbutton of the holder handle to pre-set and deliver the coagulating current (the
choice between FORCED COAG, SOFT COAG or BIPOLAR must be done pressing the
correspondent pushbutton on the unit).