Gima DIATERMO MB122 Скачать руководство пользователя страница 12

GIMA SpA

 

    DIATERMO 

MB122-MB132-MB160 

Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones 

MA036IGBEa  

 

31

 

3. INSTALLATION 

 

Inspect the unit for damages during transport. The claims for possible damages will be accepted 
only in case they are immediately communicated to the carrier; the damages that are found must 
be written down and presented to LED SpA or to your own retailer. If the unit is returned to the 
LED SpA or to your own retailer, it is necessary to use the original equipment’s package or 
another equivalent one, to guarantee the safety during the transport.  

 

Unpack the equipment and carefully study the documentation and operating instruction supplied. 
Mains voltage, indicated above the inlet, must agree with the local mains voltage (mains voltage 
frequency: 50-60 Hz). The correct voltage setting is selected by turning the voltage selector when 
available. Insert the correct fuses in the module referring to the value written on the label. 

 

Connect mains cable to a mains outlet having good hearth connection 

OPERATION OF THE EQUIPMENT WITHOUT EARTH CONNECTION IS 

FORBIDDEN 

 

The unit must be installed on a level surface, with dimension, at least, correspondent to those of 
the base of the unit itself. Around the unit must be left a space of 25cm, at least.  

 

Connect the mains cable to the mains socket on the rear panel of the unit. 

 

Connect the equipotential binding post located at the left of the unit’s back panel to equipotential 
socket of the plant. 

 

Connect the single foot switch or the double foot switch (optional) to the connector on the rear 
panel. 

 

Connect handle, in the case of use of handle without finger switch it shall be connected on the 
black buckle.  

 

In case of use of bipolar forceps (see BIPOLAR operation paragraph 4.4.5) it is necessary to use 
the special optional adapter (

REF

 00498.04). 

 

Let unit work in dry environment only. Any verified condensate must be let evaporate before 
putting in operation the unit. Don't exceed the temperature environment or the allowed moisture. 
Environmental conditions: Temperature: 10/40°C - Relative moisture: 30/75% - Pressure: 

70/106k 

Pa

 

 

When the unit is switched on, through the on/off switch on the frontal panel, after having checked 
the internal parameters, it will work with the function and the power level utilized during the last 
switching (when the unit is switched for the first time the level will be 00).  

 

Before using the unit, it is necessary connect the cable to the patient plate. Single plate electrodes 
and split plate electrodes can be. If the value of the impedance is acceptable, the OC indicator light 
will stop flashing and the alarm to sound. 

 

Having: 

Holder-handle with two pushbuttons without foot switch

: press the yellow pushbutton on the holder-

handle to deliver the cutting current (the choice between CUT or BLEND must be done pressing the 
correspondent pushbutton on the unit); or the blue pushbutton on the holder handle to deliver 
coagulating current (the choice between FORCED COAG, SOFT COAG or BIPOLAR must be done 
pressing the correspondent pushbutton on the unit). 

Holder handle with two pushbuttons and a single foot switch

: choose the cutting current CUT or 

BLEND and the coagulation current FORCED COAG, SOFT COAG or BIPOLAR. Preset through the 
yellow pushbutton on the holder handle, the function for the cut that appears on the unit or, through the 
blue pushbutton on the holder handle, the function for the coagulation that appears on the unit. The 
current delivery takes place through the foot switch.  

Holder handle with two pushbuttons and double foot switch

: press the yellow foot switch or the 

yellow pushbutton of the holder handle to pre-set and deliver the cutting current (the choice between 
CUT or BLEND must be done pressing the correspondent pushbutton on the unit) or the blue foot 
switch or the blue pushbutton of the holder handle to pre-set and deliver the coagulating current (the 
choice between FORCED COAG, SOFT COAG or BIPOLAR must be done pressing the 
correspondent pushbutton on the unit). 

Содержание DIATERMO MB122

Страница 1: ...MANUALE D ISTRUZIONI INSTRUCTION S MANUAL MANUAL DE INSTRUCCIONES DIATERMO MB122 MB132 MB160 M30540 MA036IGBEa...

Страница 2: ...rsonnel at all times All the safety instructions and advice notes are to be observed Be sure that these operating instructions is furnished together the equipment when this is transferred to other ope...

Страница 3: ...D COAG 18 4 4 4 Corrente per Coagulazione Profonda SOFT COAG 18 4 4 5 Corrente di Coagulazione Bipolare BIPOLAR 18 4 5 SEGNALAZIONE DI TEMPO DI EROGAZIONE ECCESSIVO OVT 19 4 6 SEGNALAZIONE DI ECCESSIV...

Страница 4: ...on Current SOFT COAG 35 4 4 5 Bipolar Coagulating Current BIPOLAR 35 4 5 SIGNALING OF EXCESSIVE TIME OF DELIVERY OVT 36 4 6 SIGNALING OF EXCESSIVE IMPEDANCE IN THE CIRCUIT OF NEUTRAL ELECTRODE OC 36 4...

Страница 5: ...de la Coagulacion BIPOLAR 53 4 5 EL SE ALAR DE LA POCA EXCESIVA DE LA SALIDA OVT 54 4 6 EL SE ALAR DE LA IMPEDANCIA EXCESIVA EN EL CIRCUITO NEUTRAL DEL ELECTRODO OC 54 4 7 AJUSTE DEL NIVEL AC STICO DE...

Страница 6: ...room It has foreseen its use in the monopolar cut cut coagulated or coagulation mode or in bipolar coagulation mode The equipment is conceived for being used in the following sectors DIATERMO descrizi...

Страница 7: ...00500 L10 Bent ball electrode 3mm L10 5pcs 5 cm 30529 00500 L10 L Bent ball electrode 3mm L10 L 5pcs 10 cm 30505 00500 L5 Bent hook electrode L5 5pcs 5 cm 30525 00500 L5 L Bent hook electrode L5 L 5pc...

Страница 8: ...effects can be precisely controlled it can be used safely without damage to the bone but since good coagulation while cutting is one of principal benefits of using electrosurgery a current with a cert...

Страница 9: ...ectioning of the biological tissue due to the passage of HF current concentrated on the tips of the bipolar forceps When the HF current is applied to the tissue between the two tips of the forceps an...

Страница 10: ...le or metallic support The use of antistatic sheets is advised Avoid the skin to skin contact for example between arm and body of the patient Insert an interface material like dry surgical gauze Moreo...

Страница 11: ...e lack of efficient connection to earth of the installation Operation without a protective earth connection is forbidden Before connect the equipment ascertain that the required voltage showed on the...

Страница 12: ...aporate before putting in operation the unit Don t exceed the temperature environment or the allowed moisture Environmental conditions Temperature 10 40 C Relative moisture 30 75 Pressure 70 106k Pa W...

Страница 13: ...done pressing the correspondent pushbutton on the unit press the blue foot switch to pre set and deliver the coagulating current the choice between FORCED COAG SOFT COAG or BIPOLAR must be done pressi...

Страница 14: ...work 4 1 1 MANUFACTURER S IDENTIFICATION DATA DIATERMO MB122 MB160 and MB132 HF electrosurgical unit are designed manufactured and tested by the LED SpA in its own laboratories in Aprilia LT Italy 4 1...

Страница 15: ...ion keyboard 5 Alarm warning for current delivery over time 6 OC open circuit alarm 7 Handle connection 8 Return plate connection MB132 1 Mains switch 2 Pulse time selector 3 Pulse repetition selector...

Страница 16: ...t when a deep coagulation together the cut is desired This current is made by sine current e suited for the cut associated to low voltage current suited for coagulation soft coag With this blending a...

Страница 17: ...indications 1 Switch on the unit through the mains switch while the CUT pushbutton is maintained pressed 2 When the unit has finished to check internal parameters on the display CUT appears the messa...

Страница 18: ...push the CONT push button to do it 4 10 2 REPETITION OF TIMED DELIVERY The timed high frequency current emission can be automatically repeated from 2 to 9 times each time one foot switch or one hand s...

Страница 19: ...ided with line fuses WARNING before switch on the unit operator has to verify that requested mains voltage corresponds to the voltage available from the electrical net 4 11 2 PEDAL CONNECTOR On the le...

Страница 20: ...0 1050 1050 15 Maximum output voltage BLEND Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage FORCED Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage SOFT Vpp on5 2k9 540 540 540 15 Maximum out...

Страница 21: ...of the supplier of each item When original reusable accessories supplied by LED S p A are applied the cleaning by using soft cleanser and sterilization through steam sterilisation at 121 C is recommen...

Страница 22: ...s properly grounded Assure that all the accessories that should be used are available and sterilised Check by disconnecting the reference electrode cable the functioning of the OC light and sound alar...

Страница 23: ...edance load 100 2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nomina...

Страница 24: ...2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nominal value Diagrams...

Страница 25: ...end of the life the present product mustn t be eliminated as urban refusal but it must be eliminated in a separated collection If the product is eliminated in unsuitable way it is possible that some p...