3.3 Start Test
Preparation before test:
• Urine test strips that matched the device;
• Urine sample that stored no more than 4 hours;
• Absorbent paper for sucking residual urine;
• Protective gloves for preventing contamination.
• After the test strip is immersed in the sample, absorb the excess sample
liquid in both sides of test paper with absorbent paper, then put the test strip
flat on the device test paper tray. Keep the top of test strip is aligned with
the top of test paper tray.
• The main interface displays test countdown, test sample number, user name
and the name of the test item.
According to the need, operate as follows:
Short press this button once, the sample number add 1; long press this
button for 2 seconds, the paper tray out of the storehouse.
Short press this button once, the sample number reduce 1; long press
this button for 2 seconds, the paper tray back to the storehouse.
Short press this button once to switch testing user; long press this
button for 2 seconds to enter the device setup interface; during the test,
short press it to exit.
Short press this button once to begin 60-seconds countdown, enter the
test after the countdown return to zero; short press it once again before
zeroing, the countdown will directly return to zero and immediately enter
the rapid test status.
Long press this button for 1 second, the device will power off.
Test steps:
1) Place the test strip;
2) Under the main interface, press OK button to begin testing the current
sample;
3) Begin 60-seconds countdown. After finishing the countdown or press OK
button once again, the device begin to test data. If there are error messages
during the test, follow the prompt, then press OK button to continue
the measurement;
4) Measurement is completed, display and store the test result. As Figure 3-3.
Note
• The sample number starting from 1 after the first power-on every day, after
testing 1 sample, the sample number automatically add 1. After power