29
Chapter 2
INSTALLATION
2.1 Operation Environment
As with all precise electronic instruments, the urine analyzer should avoid
placing in high temperature and humidity environment for a long time.
To get optimal result, please keep relative stable temperature (10°C~30°C) and
humidity (≤80%), and the tabletop to be placed the analyzer should be level.
Operating environment:
Temperature: 10°C~30°C
Relative humidity: ≤80%
Atmospheric pressure: 76kPa~106kPa
Transportation and storage environment:
Temperature: -40°C~55°C
Relative humidity: ≤95%
Atmospheric pressure: 76kPa~106kPa
Warning
Please don’t use the analyzer in the following places:
• Direct sunlight areas or the front of open window;
• There is flammable and explosive gases;
• Near the heating or cooling equipment;
• Near strong light-source.
2.2 Dismantlement and Installation
Open the package and remove the material used for transportation. Keep the
package for possible future transportation or storage.
1) Take out the urine analyzer from the package.
2) Remove the wrapper, take out the analyzer from the plastic packaging.
3) Check the components according to the packing list.
If there is any problem, contact our company or agent immediately.