![Gima 32916 Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/gima/32916/32916_manual_2227740057.webp)
57
Guidance and manufacturer’s declaration - electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test specifications
for ENCLOSURE
PORT IMMUNITY
to RF wireless
communications
equipment
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Modulation
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385
380-390 TETRA
400
Pulse
modulation b)
18Hz
1.8
0.3
27
450
430-470 GMRS 460,
FRS 460
FM c) ± 5kHz
deviation 1kHz
sine
2
0.3
28
710
704-787
LTE Band
13,
17
Pulse
modulation b)
217Hz
0.2
0.3
9
745
780
810
800-960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18Hz
2
0.3
28
870
930
1720
1700-
1990
GSM 1800;
CDMA
1900;
GSM 1900;
DECT; LTE
Band 1, 3,
4, 25;
UMTS
Pulse
modulation b)
217Hz
2
0.3
28
1845
1970
2450
2400-
2570
Bluetooth,
WLAN,
802.11
b/g/n, RFID
2450, LTE
Band 7
Pulse
modulation
217 Hz
2
0.3
28
5240
5100-
5800
WLAN
802.11
a/n
Pulse
modulation
217 Hz
0.2
0.3
9
5500
5785
Table 3
Disposal: The product must not be disposed of along with other domestic waste. The
users must dispose of this equipment by bringing it to a specific recycling point for
electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
ENGLISH