background image

21

Filter assembling

Make sure the filter is assembled with the arrows on the side of the patient.

WARNING: The inside of the medical device must be regularly checked for the presence of liquids 

or other visible contamination (secretions). In the presence of liquids or other visible contamination, 

immediately replace the medical device due to the risk of an insufficient vacuum flow rate.

These products have been designed, tested and manufactured exclusively for single patient use and 

for a period no longer than 24 hours.

Operation using cigarette lighter DC 12V

•  Connect the device’s external plug 12V to the ligther plug with the cigarette lighter cable. Check the battery 

power status of the vehicle before the cigarette lighter cable. Press the switch to start suction

•  Press the switch to the I position to turn it on.

WARNING: Only use the originally supplied or recommended replacement cigarette ligther cables 

(view the chapter “Important Safety Rules”)

Operation with Internal Battery 

•  Press the switch in position I to turn the device on (the external power pack doesn’t need to be connected)

•  The fully charged battery life is about 60 minutes with continuous operation.

 WARNING: Before using the device, check the battery power status. Before each use proceed with 

charging the battery. To maintain the device in good conditions, recharge the battery every 3 months (when 

not in use).

Recharging operations: 

to be able to charge the internal battery it is necessary to connect the universal 

switching adapter to the electric network for approx. 240 minutes with the main switch to position 0. 

ENGLISH

Содержание 28243

Страница 1: ...RADOR SUPER VEGA BATTERY ASPIRADOR SUPER VEGA BATTERY SSAK AKUMULATOROWY SUPER VEGA BATTERY ASPIRATOR SUPER VEGA BATTERY SUPER VEGA BATTERY M28243 M Rev 2 09 20 PROFESSIONAL MEDICAL PRODUCTS Gima S p...

Страница 2: ...ilst the plug is still connected disconnect the mains switch remove the plug from the power supply and contact the GIMA technical service department Do not attempt to make the device work before it ha...

Страница 3: ...in Home Care Keep all accessories of the device out of reach of children under 36 months of age since they contain small parts that may be swallowed 15 Do not leave the device unattended in places ac...

Страница 4: ...rotecting patient and machine against infections cross It blocks the passage of the liquids that come into contact with it Always replace it if yes suspected to be contaminated and or wetted or discol...

Страница 5: ...part of the container from secretions under cold running water and then clean every single part in hot water temperature not exceeding 60 C Once again carefully wash each single part using if necessar...

Страница 6: ...order to replace it consult the technical staff authorised by the manufacturer USE ONLY THE RECOMMENDED BATTERIES FROM GIMA THE USE OF OTHER BATTERIES ARE NOT RECOMMENDED AND INVOLVING THE CANCELLATIO...

Страница 7: ...nd must be connected to the spout on the vase lid with the float assembled inside over flow device The overflow device is triggered the float closes off the internal lid fitting when the maximum volum...

Страница 8: ...hicle before the cigarette lighter cable Press the switch to start suction Press the switch to the I position to turn it on WARNING Only use the originally supplied or recommended replacement cigarett...

Страница 9: ...lose proximity with adjacent to or on top of the medical device If such use is necessary and unavoidable special precautions should be taken so that the electro medical device functions properly in it...

Страница 10: ...The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it s u...

Страница 11: ...rmined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified b...

Страница 12: ...ransmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequency where P is...

Страница 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Отзывы: