21
Filter assembling
Make sure the filter is assembled with the arrows on the side of the patient.
WARNING: The inside of the medical device must be regularly checked for the presence of liquids
or other visible contamination (secretions). In the presence of liquids or other visible contamination,
immediately replace the medical device due to the risk of an insufficient vacuum flow rate.
These products have been designed, tested and manufactured exclusively for single patient use and
for a period no longer than 24 hours.
Operation using cigarette lighter DC 12V
• Connect the device’s external plug 12V to the ligther plug with the cigarette lighter cable. Check the battery
power status of the vehicle before the cigarette lighter cable. Press the switch to start suction
• Press the switch to the I position to turn it on.
WARNING: Only use the originally supplied or recommended replacement cigarette ligther cables
(view the chapter “Important Safety Rules”)
Operation with Internal Battery
• Press the switch in position I to turn the device on (the external power pack doesn’t need to be connected)
• The fully charged battery life is about 60 minutes with continuous operation.
WARNING: Before using the device, check the battery power status. Before each use proceed with
charging the battery. To maintain the device in good conditions, recharge the battery every 3 months (when
not in use).
Recharging operations:
to be able to charge the internal battery it is necessary to connect the universal
switching adapter to the electric network for approx. 240 minutes with the main switch to position 0.