23
or malfunctions). The use of accessories, transducers and cables differing from those specified, with the ex-
ception of transducers and cables sold by the appliance and system manufacturer as spare parts, can lead to
an increase in emissions or in a decrease of the immunity of the device or system. The following tables supply
information regarding the EMC (Electromagnetic Compatibility) characteristics of the electro-medical device.
ENGLISH
Guidance and manufacturer’s declaration – electromagnetic Emissions
The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it’s used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
Irradiated / Conducted
emissions CISPR11
Group 1
The SUPER VEGA BATTERY SUCTION UNIT only used RF energy
only for its internal functioning. Therefore, its RF emissions are very
low and are not cause interference in proximity of any Electronic ap-
pliances.
Irradiated / Conducted
emissions CISPR11
Class [B]
The SUPER VEGA BATTERY SUCTION UNIT can be used in all
environments, including domestic and those connected directly to the
public mains distribution that supplies power to environments used
for domestic scopes.
Harmonic emissions
EN 61000-3-2
Class [A]
Voltage fluctuations / flicker
emissions EN 61000-3-3
Complies
Guidance and manufacturer’s declaration – Immunity Emissions
The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it’s used in such an
environment.
Immunity Test
Level indicated by the
EN 60601-1-2
Compliance Level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
+/-8kV on contact
+/-15kV in air
The device doesn’t
change its state
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic ma-
terial, the relative humidity should be at least
30%.
Electrical fast
transient / burst
EN 61000-4-4
± 2kV power supply
lines
± 1kV for input /
output lines
The device doesn’t
change its state
Mains power quality should be that of a typical
commercial environment or hospital.
Surge
EN 61000-4-5
± 1kV differential mode
+/-2 kV ordinary mode
The device doesn’t
change its state
Mains power quality should be that of a typical
commercial environment or hospital.
Loss of voltage, brief
voltage interruptions
and variations
EN 61000-4-11
<5% U
T
(>95% dip U
T
)
for 0,5 cycle
40 % U
T
(60% dip U
T
)
for 5 cycle
70 % U
T
(30% dip UT )
for 25 cycle
<5 % U
T
(>95% dip U
T
)
for 5 sec
- -
Mains power quality should be that of a typi-
cal commercial environment or hospital If the
user of the SUPER VEGA BATTERY SUC-
TION UNIT request that the appliance oper-
ates continuously, the use of a continuity unit
is recommended.
Magnetic field with
network frequency
(50/60 HZ)
EN 61000-4-8
30A/m
The device doesn’t
change its state
The power frequency magnetic field should
be measured in the intended installation lo-
cation to make sure that it’s sufficiently low.
Note U
T
is the value of the power supply voltage