   GB  

17

 

Guidance and manufacturer’s declaration – Immunity Emissions 

The 

EOLO

 Nebulizer

 

is intended for use in the electromagnetic environment specified below.

The customers or the user of the 

EOLO

 Nebulizer

 

should assure that it’s used in such an environment. 

Immunity Test 

Level indicated by the 

EN 60601-1-2 

Compliance level 

Electromagnetic environments - 

guidance 

Conducted Immunity 
EN 61000-4-6 

3Vrms 150kHz to 80Mhz  

(for non life-supporting 

devices) 

V

1

 = 3 V rms 

The portable and mobile RF communication 
devices, including cables, must not be used 

closer to the EOLO device, than the 
separation distance calculated by the 
equation applicable to the transmitter 

frequency. Recommended separation 
distance 

d = [3.5 / V

1

P

d = [12 / E

1

P

from 80 MHz to 800MHz

d = [23 / E

1

P

from 800 MHz to 2.5 GHz

Where P is the maximum nominal output 
voltage of the transmitter in Watt (W) 

depending on the manufacturer of the 
transmitter and the recommended 
separation distance in metres (m). The 

intensity of the field from the fixed RF 
transmitters, as determined by an electro-
magnetic study of the site

a)

,  could be lower 

than the level of conformity of each 
frequency interval 

b)

. It is possible to check 

for interference in proximity to devices 
identified by the following symbol: 

Radiated Immunity 

EN 61000-4-3 

3V/m 80MHz to 2.5GHz 

(for non life-supporting 

devices) 

E

1

 = 3 V / m 

Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied 
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the 
absorption and by reflection from buildings, objects and people. 

a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terrestrial

mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and 
accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnetic 

study of the site should be considered. If the field intensity measured in the place where the device will be used surpasses the 
above mentioned applicable level of conformity, the normal functioning of the device should be monitored.  If abnormal 

performance arises, additional measures such as changing the device's direction or positioning may be necessary. 

b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.

Recommended separation distance between portable and mobile radio-communication devices and the monitor 

The EOLO

 Nebulizer

 is intended to operate in an electro-magnetic environment where RF irradiated interferences are 

under control.  The client or operator of the EOLO device can help prevent electro-magnetic interference by keeping a 
minimum distance between the portable and mobile RF communication devices (transmitters) and the EOLO device, as 
recommended below, in relation to the radio-communication maximum output power. 

Maximum nominal output 

power  of the 

Transmitter W 

Separation distance

 

from the frequency transmitter (m) 

150 kHz to 80 MHz 

d = [3.5 / V

1

P

80 MHz to 800 MHz 

d = [12/E

1

P

800 MHz to 2.5 GHz 

d = [23/E

1

P

0,01 

0,12 

0,12 

0,23 

0,1 

0,38 

0,38 

0,73 

1,2 

1,2 

2,3 

10 

3,8 

3,8 

7,3 

100 

12 

12 

23 

For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres 
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output 

power of the transmitter in Watt (W) depending on the transmitter's manufacturer. 

Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied 

Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the 
absorption and by the reflection from buildings, objects and people. 

Содержание 28097

Страница 1: ...1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy Made in Italy Instruction manual EOLO...

Страница 2: ...vice in its own box away from dust and sunlight In general it is inadvisable to use single or multiple adapters and or extensions Should their use be necessary you must use ones that are in compliance...

Страница 3: ...110 kPa Neb Rate with 4ml of 0 9 NaCl solution 0 40 ml min with 4ml of 0 9 NaCl solution MMAD 2 44 m GSD 2 87 Weight 1 65 Kg Size 148 x 223 h x 124 h Noise Level measured as specifications of EN 13544...

Страница 4: ...0 OFF Lot Number Serial Number Identification device IP21 Degree of protection an electrical device provides in the case of accidental or intentional contact with the human body or with objects and pr...

Страница 5: ...size to contain all the individual components with a solution of drinking water and disinfectant hypochlorite based solution readily available in a pharmacy by following the proportions indicated on...

Страница 6: ...e lid Connect air pipe 5 to the air exit well 4 Connect the other end of the pipe to the connection in the lower part of the nebulizer Connect the desired accessory to the nebulizer child mask or adul...

Страница 7: ...ny Electronic appliances Irradiated Conducted emissions CISPR11 Class B The EOLO Nebulizer can be used in all environments including domestic and those connected directly to the public mains distribut...

Страница 8: ...TV transmitters can not be theoretically and accurately foreseen To establish an electro magnetic environment generated by fixed RF transmitters an electro magnetic study of the site should be conside...

Страница 9: ...have been checked we suggest to contact your dealer or technical service GIMA BEFORE EVERY CHECKING OPERATION IN CASE OF ANOMALIES OR BAD FUNCTIONING PLEASE CONTACT GIMA TECHNICAL SERVICE GIMA DOES N...

Страница 10: ...Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy...

Отзывы: