TROUBLESHOOTING
|
PIPETMAN® CLASSIC
Chapter 5
TROUBLESHOOTING
A quick inspection of the pipette may help you to
detect a problem.
NOTE
You may download the Two Minute
Inspection Poster from the Gilson
website (
), which
shows how to perform a quick
diagnosis of your pipette.
WARNING
Before returning any pipette to
your local Gilson Service Center,
ensure that it is completely free of
chemical, biological, or radioactive
contamination (refer to
on page 18).
Please use the included safety bag
to return the pipette to your local
Gilson Service Center.
Troubleshooting Table
The following table may help you to identify and
correct issues you might encounter.
SYMPTOM
POSSIBLE CAUSE
ACTION
Pipette is
leaking sample
Damaged tip holder
Replace the tip
holder
Worn O-ring or seal
Replace both parts
Pipette won’t
aspirate
Worn O-ring or seal
Replace both parts
Damaged tip holder
Replace the tip
holder
Connecting nut is
loose
Tighten connecting
nut
Damaged or corroded
piston
Return pipette to
supplier
Improper repair or
assembly
Refer to page 15
Pipette is
inaccurate
Improper repair or
assembly
Refer to page 15
Unscrew tip holder
Tighten connecting
nut
Connecting nut is
loose
Tighten connecting
nut
Содержание PIPETMAN Classic P10
Страница 1: ...User s Guide EN CLASSIC ...
Страница 25: ...Notes 25 PIPETMAN Classic USER S GUIDE NOTES ...
Страница 26: ...Notes 26 Notes PIPETMAN Classic NOTES ...
Страница 27: ...Notes 27 PIPETMAN Classic USER S GUIDE NOTES ...
Страница 28: ...www gilson com contactus LT801120 H 2019 Gilson Inc All rights reserved ...