MAINTENANCE
|
PIPETMAN® CLASSIC
Changing the O-Ring
The O-ring and seal on the piston must not be
autoclaved. If worn or damaged in any way
(chemical or mechanical), they must be replaced.
The dimensions of the O-ring vary depending on
the model of pipette.
1.
Remove the tip ejector (refer to
2. Unscrew the connecting nut by turning it
counterclockwise.
3. Carefully separate the lower and upper parts.
4. Remove the piston assembly, O-ring, and
seal.
5. Clean or replace the seal followed by the O-ring.
6. Reassemble the pipette (refer to Figure 7 on
page 16).
7. Tighten the connecting nut (turn clockwise).
8. Refit the tip ejector (refer to
Cleaning and Decontamination
PIPETMAN Classic is designed so that the parts
normally in contact with liquid contaminants
can easily be cleaned and decontaminated.
However, because the models P2 and P10 contain
miniaturized parts, it is best not to disassemble
these pipettes yourself. Please contact your
authorized Gilson Service Center.
NOTICE
You can refer to the
decontamination procedure
available on the Gilson website
). Liquid must
never enter the upper part
(handle) of any pipette.