9
MyPIPETMAN®
|
QUICK GUIDE
QUICK GUIDE
| MyPIPETMAN®
Multichannel Models
PIPETMAN DIAMOND Tips: TIPACK AND TOWERPACK
PIPETMAN DIAMOND tips are best fitted with the
ROCKY RACK technique, invented by Gilson, available
only in our TIPACK and TOWERPACK.
PIPETMAN DIAMOND Tip
Compatibility for Multichannel Models
P8x10, P12x10
D10*, DL10, DF10ST,
DFL10ST
P8x20, P12x20
DL10, D200, DFL10ST,
DF30ST
P8x200, P12x200
D200, D300,
DF200ST, DF300ST
P8x300, P12x300
D200, D300,
DF200ST, DF300ST
*Using an ejector adapter, you can adapt a
D10 tip (refer to page 8).
Pre-Wetting the Tips
Pre-wetting the tips before pipetting helps prepare the tips
for the best pipetting performance. Ideally, the pre-wet
includes both immersing the tip in the liquid and performing
one pipetting step.
Pre-wetting the tips helps ensure that volumes that you pipette
will achieve accuracy and precision within specifications.
Aspirating
1.
Press the push button to the
first stop
(this corresponds
to the set volume of liquid).
2.
Hold the pipette vertically and immerse the tip in the
liquid.
3.
Release the push button slowly and smoothly to the top
position to aspirate the set volume of liquid.
4.
Wait one second and then withdraw the pipette tip from
the liquid. You may wipe any droplets away from the
outside of the tip using a medical wipe. However, if you do
so, take care to avoid touching the tip’s orifice.
Figure 6
ROCKY RACK
technique
Содержание MyPipetman P10