6
6 - FIttIng the caPILLary and PIston
Verify that the model of capillary-
piston (CP) corresponds to the
MiCroMan you are preparing for use:
cP25, cP50 and cP250 Bulk
Capillaries and pistons are delivered in
two separate boxes. To fit a capillary-
piston, follow this procedure:
1) Take care not to damage the sealing tip
1
when you
handle the piston
3
.
2) Press the push-button to the second stop to open
the clamp
5
.
6
5
MicroMan
cP
model
CP10
m10
CP25
m25
CP50
m50
CP100
m100
CP250
m250
CP1000
m1000
2
3
4
1
Never lubricate the capillary-holder, capillary,
or any other part of the pipette. If the capillary
tends to slip off the capillary-holder, clean the
capillary-holder with ethanol, using a medical wipe
or similar soft tissue.