9
To avoid parallax
errors, make sure that
the volume indicator
and the selected
volume marking are
in your direct line of
vision. at close range
you may find it helps
to close one eye.
8 - PIPettIng
aspirating
- press the push-button to the
first stop (1),
- immerse the capillary 2 mm
into the liquid,
-
slowly, release the push-
button to draw up the
liquid (top position),
- wipe any liquid from the out-
side of the capillary, taking
care not to touch the orifice.
dispensing
- place the end of the capillary against the inside wall
of the recipient vessel,
- press the push-button slowly to the first stop,
- keeping the push-button depressed, move the capillary
away from the side wall,
- withdraw MiCroMan from the vessel and release the
push-button.
ejecting the Piston-capillary:
Press the push-button to the first stop (1), then press
harder to the second stop (2); at this point the piston
and capillary are ejected simultaneously.
Top
1
2