31
ASPEC® POSITIVE PRESSURE MANIFOLD
|
USER'S GUIDE
M
AINTENANCE
Maintenance
Clean the ASPEC PPM
The instrument should be cleaned occasionally, using a dry, clean cloth. If
necessary, use a cloth dipped in soapy water. If liquid is accidentally spilled
on the instrument, wipe it up immediately.
Install or Replace the O-Ring
1.
Remove the manifold from the manifold carrier.
2. Gently remove the O-ring from the rear of the manifold with your
fingers. If necessary, free the O-ring with a flathead screwdriver.
3. Insert a new O-ring into the circular depression in the rear of the
manifold.
Replace the Manifold Gasket
1.
Remove the manifold from the manifold carrier.
2. Set the manifold down with the gasket facing up.
3. Peel the gasket away, starting from the corner. Discard the gasket.
4. Clean off any excess adhesive from the manifold using methanol.
5. Locate the replacement gasket.
6. Peel away the adhesive backing on the new gasket, starting from the
corner. Discard the backing material.
7. Align the new gasket with the four corners of the manifold before
attaching the gasket with the adhesive side facing down.
Maintenance Parts
Description
Part Number
Gasket, 96-position for SPE manifold (qty. 10)
37417005
Gasket, 24-position for SPE manifold (qty. 10)
37417006
Gasket, 15-position for SPE manifold (qty. 10)
37417007
O-ring for manifold seal (qty. 5)
37417008