23
ASPEC® POSITIVE PRESSURE MANIFOLD
|
USER'S GUIDE
O
PERA
TION
Setup for Flow to Collection
96-Position
1.
Slide the locator plate to
the front of the locator bed.
2. Move the drain well to the
rear position of the locator
plate.
3. Place the 96-position well
plate on the front position
of the locator plate.
4. Set the plate spacer on the
96-position well plate.
5. Set the appropriate SPE
plate on the plate spacer.
15- or 24-Position
1.
Slide the locator plate to
the front of the locator bed.
2. Move the drain well to the rear
position of the locator plate.
3. Place the 15- or 24-position
tube rack with tubes on the
front position of the locator
plate.
4. Set the appropriate 15- or 24-position SPE cartridge rack on the tube
rack assembly.
NOTE
If your experiment does not require the full use of the 15- or
24-position rack, use plugs to fill the empty spaces. Refer to
SPE Cartridge Plugs (Optional)
on page 15.
Figure 17
Flow to Collection Setup (96-Position)
96-Position
Well Plate
Locator
Plate
SPE Cartridge Rack
(96-Position)
Plate Spacer
Collection and
Drain Well
Front
Rear
Figure 18
Flow to Collection Setup (24-Position)
(15-Position SPE Rack Not Shown)
Collection
Tubes and
Tube Rack
SPE Cartridge
Rack
(24-Position)