Leveling the Base Assembly
Upon completion of the above adjust the 3 leveling thumb nuts under the tipping
bucket to align the spirit level bubble to within the centre circle.
Now tighten the upper 3 thumb nuts ensuring that the spirit level bubble remains
within the centre circle.
NOTES:
Ensure that the Foam insert under the tipping bucket is removed before re-fitting
the funnel.
No two rain gauge designs are ever likely to produce identical results, and identical
rain gauge can give slightly different catches even when sited close to each other.
Customers with 2 or more sets of equipment must not swap around rain gauges
with MetPaks without using MetSet to change the calibration figure.