2.2
Installation
2.2.1
Bench System Test
Prior to physically mounting the MetPak RG in its final location, we strongly
recommend that a bench system test is carried out to confirm the system is
configured correctly, is fully functional and electrically compatible with the selected
host system and cabling (preferably utilising the final cable length). The required
data format, units, output rate, and other options should also all be set up at this
stage.
Note that MetPak RG is supplied with a matching Rain Gauge and its tip calibration
factors will have been pre-programmed into the MetPak hub box. If a replacement
Rain gauge is used then the Hub box will need re-programming using Gill MetSet
Software with the new tip calibration factor (if this has changed).
2.2.2
Bench Test Equipment Required
1723-3B-2-111 MetPak RG set for RS232 communication (factory default setting)
and matching Rain Gauge.
1723-10-051
MetPak RG RS232 to USB 1.8M configuration cable with integral
5v power or alternatively use RS232, RS232 converter and
separate 5v to 30 dc power supply.
ES-U-1001-A EasySynch RS232 to USB adaptor or equivalent (if not using Gill
part 1723-10-051) or connect directly to an RS232 COM port.
Power Supply 5V to 30Vdc at 100mA if required.
Belden 9503 3 pair communications and power cable (if not using Gill Part 1723-
10-051), length as required.
Digital Multimeter capable of measuring resistance and dc volts.
Gill MetSet and MetView Software supplied on the MetPak RG CD or download
from.
http://www.gill.co.uk/main/software.html
2.2.3
Cabling
MetPak RG has five communication connection options:
USB (using the 1.8m Gill USB cable, Part No. 1723-10-051).
RS232
RS422
RS485 (two wire point to point).
SDI-12
MetPak RG has a Rain Gauge Digital Input connection:
Digital Switch (DIG1)
It is important that the cable is appropriate for the chosen connection option. The
following sections describe the recommended types and maximum lengths of cable
in each case.