IOM Sampling with the Gilian GilAir Plus Pump Procedure
7
4. Run tubing under the arm to collar/breathing zone area. Clip IOM on shirt.
Make sure IOM filter is open to the front (to allow proper air flow) and is
stationary.
5.
Use additional clips on tubing to attach to the person’s shirt, allowing a little slack
on tubing as shown in photo 8.
6. Remove cap from the IOM.
Photo 8. IOM Filter Correctly Oriented on Wearer, with Cap Removed
7. Turn pump on by pressing the check mark/power button with the indicator on the
“Run” option.
The flow will gradually increase to 2000 cc/min. A flashing green light will
appear at the top of screen. This indicates that the pump is in fact in the
run mode.