Operation
Manual 20791, Rev. C, June 2001
2-99
The [
CW/SWEPT
] softkey in the DEFINE menu, and the [
GRAPH/READOUT
] softkey in the
DISPLAY menu give you complete flexibility in configuring the analyzer measurement modes and
display. The [
ALL SWEPT
] and [
ALL CW
] softkeys in the MEAS menu combine the functions to make
switching between modes much faster and simpler.
2.12.3
Cal Factor Corrections in the CW Mode
When a channel is set to CW mode, the readings are corrected for Cal Factor. The 8003 Analyzer
assumes that the signals measured are all at the same frequency that the source is set to.
Sometimes the sensors measure at a frequency different than that of the source. This may happen if
there is frequency translation (mixers or up/downconverters) or frequency multiplication/division. In
this case, you can manually enter the correct frequency by pressing
[CONFIG]
followed by the
[
SENSORS CW FREQ
] softkey. You will see that the [
FROM SOURCE
] softkey is in light blue -
indicating that the analyzer is currently using the CW frequency it commanded the source to go to.
Press the [
CW FREQ—
] softkey and the appropriate [
X CW FREQ
] softkey where X is the sensor letter.
Then enter your desired operating frequency on the numeric entry keypad followed by the
[GHz/dB]
units key (see Figure 2-56).
Another way to make a Cal Factor correction is by direct entry of Cal Factor. This may be necessary if
the sensor is user calibrated (perhaps with an adapter), and the user Cal Factors are not entered in the
sensor EEPROM table. To make a direct Cal Factor correction, press the
[CONFIG]
key followed by
the [
SENSORS CW FREQ
] softkey. Press the [
CW CAL FACTOR
] softkey, then the appropriate [
X
CAL FACTOR
] softkey where X is the sensor letter.
NOTE:
A Power Sweep Measurement capability is also available with the 8003 instrument.
This feature is described in the Power Sweep Measurements Section 2.10.
Figure 2-56: 3 Different Methods (CAL Factor Correction Plus Sensor Offsets)
☛
☛
☛
☛
SENSOR
A OFFST
CLR ALL
OFFSETS
CONFIG
SEN A
OFFS
0.00 dB
FREQ
18.00 GHz
SEN B
OFFS
0.00 dB
FREQ
8.40 GHz
SEN C
OFFS
20.00 dB
CALF
-1.20 dB
RAT C/A
OFS
20.00 dB
CF C
-1.20 dB
∆
-11.39 dBm
-33.37 dBm
18.39 dBm
29.79 dBr
SENSOR
B OFFST
SENSOR
C OFFST
RETURN
CH1 - SPECIAL CW FREQ
CH2 - DEFAULT SOURCE FREQ
CH3 - SPECIAL CAL FACTOR
AND OFFSET
SEN C OFFSET
20.00 dB
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...