page 8 of 14
Operating Manual RS MH 3161-
… - ex
H61.0.RS.6C-04
5 Start Operation
Connect sensor, turn on device via
key.
After segment test
the device displays some configuration:
• If a
zero point adjustment
was carried out the display shows shortly „nuLL Corr“.
After that the device is ready for measuring.
6 Configuration
To change device settings, press
Menu
(key 4) for 2 seconds. This will call the configuration menu.
Pressing key
Menu
jumps between the parameters.
The parameters can be changed with
(key 2) or
(key 5).
Quit
(key 6) finishes the configuration and returns to standard measuring operation.
Parameter
Value
Description
‚Menu‘
or
mbar, bar, …
Unit
: Unit of display
1 … 120
Auto Power Off
time in minutes
oFF
Auto Power Off
deactivated
01, 11 … 91
Base
address
of interface
OFFS
refer to list below
The
offset of sensor
will be displaced by this value to compensate for
deviations in the probe or in the measuring device.
oFF
Zero displacement inactive (=0.00)
SCAL
-2.000 … 2.000
The measuring
scale of sensor
will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device
oFF
Scale correction factor inactive (=0.000)
Device type
adjustable offset
RS MH 3161-002 - ex
-50.0 ... 50.0 Pa
RS MH 3161-13 - ex
-500 ... 500 mbar
7 Remarks To Special Features
7.1 Power off Time
If there won’t be pressed any key and no interface communication takes place for the time of the power off
time setting (P.Off), the device will be switched off automatically to save battery power.
If P.oFF = oFF then the automatic switch off is deactivated.