2.3  Safety guidelines 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety 
measures and special safety advises given in this manual will be adhered to when using the device. 

1. 

 

Consider the operating instructions and the regulations referring the use of electrical 
equipment for hazardous areas (e.g. VDE0165) 

2. 

 

The operation in Ex-Zone is only allowed in the accompanying leather case! 

3. 

 

Only the usage of approved batteries is allowed! 

Battery exchange must only be made outside of the hazardous area! 

Approved batteries are: 

Battery type  Manufacturer 

Battery name 

6F22 

GP 

GREENCELL , 9V  (1604G) 

6LF22 

or

 

6LR61 

or

 

6LP3146 

GP 

SUPER Alkaline, 9V  (1604A) 

Duracell 

DURACELL PLUS, Alkaline, 9V 

Varta 

power

one alkaline, 9V  (No. 4122) 

Varta 

INDUSTRIAL, Alkaline, 9V  (No. 4022) 

 

4. 

 

Mains operation: 

only use power supply’s of the type GNG 10/3000! 

The operation with external power supply is not allowed in ex Protection-Zone! 

5. 

 

Only use interface converters of the type GRS 3100, USB 3100, USB 3100 N and GRS 3105! 

The operation of serial interface is not allowed in Ex Protection-Zone. 

6. 

 

Potential equalisation: 

All components (pressure sensor, power supply unit, interface, etc.) connected to the device 
must be on the same potentials! If this is not guaranteed, you have to connect them for a 
potential equalisation. 

7. 

 

Ambient conditions:

 

 

Take care that the device is not exposed to environments that make the intrusion of humidity, 
water, conducting liquids or dust possible. 

8. 

DANGER

 

Trouble-free operation and reliability of the device can only be guaranteed if the device is not 
subjected to any other climatic conditions than those stated under "Specification". 
If the device is transported from a cold to a warm environment condensation may cause in a 
failure of the function. In such a case make sure the device temperature has adjusted to the 
ambient temperature before trying a new start-up. 

9. 

DANGER

 

If there is a risk whatsoever involved in running it, the device has to be switched off 
immediately and to be marked accordingly to avoid re-starting.  
Operator safety may be a risk if:. 
   - there is visible damage to the device. 
   - the device is not working as specified. 
   - the device has been stored under unsuitable conditions for a longer time. 

In case of doubt, please return device to manufacturer for repair or maintenance. 

10. 

DANGER

 

When connecting the device to other devices the connection has to be designed most 
thoroughly as internal connections in third-party devices (e.g. connection GND with protective 
earth) may lead to undesired voltage potentials that can lead to malfunctions or destroying of 
the device and the connected devices. 

Содержание GREISINGER RS PRO 205-1444

Страница 1: ...H61 0 RS 6C 04 Operating Manual EN RS MH 3161 002 ex RS Stock number 205 1446 RS MH 3161 13 ex RS Stock number 205 1444 device without case device in leather case ready for use in Ex Zone rspro com ...

Страница 2: ...3 CONNECTIONS 7 4 4 POP UP CLIP 7 5 START OPERATION 8 6 CONFIGURATION 8 7 REMARKS TO SPECIAL FEATURES 8 7 1 POWER OFF TIME 8 8 OUTPUT 9 8 1 SERIAL INTERFACE 9 9 INPUT ADJUSTMENT 9 9 1 ZERO DISPLACEMENT SENSOR OFFS 9 9 2 SCALE CORRECTION SENSOR SCAL 9 10 PRESSURE CONNECTION 10 10 1 DEVICE TYPE WITH RELATIVE PRESSURE 10 11 ERROR AND SYSTEM MESSAGES 10 12 SPECIFICATION 11 13 RESHIPMENT AND DISPOSAL 1...

Страница 3: ...of the device misuse or malfunction of the connection or of the device The manufacturer is not liable for misprints 2 Safety 2 1 Intended Use The safety requirements see below have to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Protect the device from ...

Страница 4: ...s pressure sensor power supply unit interface etc connected to the device must be on the same potentials If this is not guaranteed you have to connect them for a potential equalisation 7 Ambient conditions Take care that the device is not exposed to environments that make the intrusion of humidity water conducting liquids or dust possible 8 DANGER Trouble free operation and reliability of the devi...

Страница 5: ... safe device again you have to check for visible damages and control functionality before you put it to the leather bag 14 Any changes or repair of the device is not allowed Please return device to manufacturer for repair or maintenance 3 Product Specification 3 1 Scope of supply The scope of supply includes Measuring device with 9V battery Operation manual Leather case 3 2 Operation and maintenan...

Страница 6: ...figuration menu Store Quit press short hold function the last measuring value will be held in the secondary display press again hides the value Please Note Activating deactivating tara clears the max min memories 1 Zero Point Adjustment If there is no pressure or zero pressure absolute applied to the pressure ports the device will display 0 If there is a permanent deviation and device is operated ...

Страница 7: ...dling Pull at label open in order to swing open the pop up clip Pull at label open again to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up...

Страница 8: ... 11 91 Base address of interface OFFS refer to list below The offset of sensor will be displaced by this value to compensate for deviations in the probe or in the measuring device oFF Zero displacement inactive 0 00 SCAL 2 000 2 000 The measuring scale of sensor will be changed by this factor to compensate deviations of temperature probe or measuring device oFF Scale correction factor inactive 0 0...

Страница 9: ...interface are always in the selected measurement unit mbar bar Supported functions Code Name Function Code Name Function 0 Read measurement value 200 Read min display range 3 Read system state 201 Read max display range 6 Read min memory 202 Read display range unit 7 Read max memory 204 Read display range decimal point 12 Read ID number 208 Read of channels 174 Clear min memory 214 Read scale adju...

Страница 10: ...000 2000 mbar 2000 0 mbar 11 Error And System Messages Display Meaning What to do Low battery power device will only continue operation for a short period of time Replace battery Battery empty Replace battery Mains operation without battery wrong voltage Check power supply replace it when necessary No display or confused characters device does not react on keypress Battery empty Replace battery Ma...

Страница 11: ...or 8 x 1 mm 6 mm inner tube Ø tubes at the top of device Display 2 four digit LCDs 12 4 mm high and 7 mm high for measuring values and for min max memories hold function etc as well as additional functional arrows Pushbuttons 6 membrane keys Output 3 5 mm audio plug stereo Interface Serial interface 3 5mm jack can be connected to USB or RS232 interface of a PC via electrically isolated interface a...

Страница 12: ...t package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 13 2 Disposal instructions Batteries must not be disposed in the regular domestic waste but at the designated collecting points The device must not be disposed in the unsorted municipal waste Send the device directly to us sufficiently stamped if...

Страница 13: ...H61 0 RS 6C 04 Operating Manual RS MH 3161 ex page 13 of 14 16 EC Type Examination Certificate ...

Страница 14: ...page 14 of 14 Operating Manual RS MH 3161 ex H61 0 RS 6C 04 17 EU Declaration of conformity ...

Отзывы: