4 Handling
4.1 Display
1
Main display:
shows actual value
2
Arrow points to the chosen measuring
unit
3
Secondary display:
shows min./max. or hold value
4
SL:
appears if sea-level-correction is activated
(only at devices with absolute pressure sensor)
5
Tara:
appears if tara-function is activated
6
Not used
4.2 Basic Operation
On / Off
+
min/max measuring:
press short:
shows the min./max. value
press again:
hides min./max. value
press 2 sec.:
clears particular value
Tara, zero-point adjustment:
press short:
display will be set to 0
The following measuring will be relatively
displayed to the set tara value
press 2 sec.:
deactivates tara-function
press 5 sec.:
Zero-Point Adjustment
1)
Set/Menu
:
press short:
invokes configuration menu
Store/Quit
:
press short:
hold-function, the last measuring value will
be held in the secondary display.
press again:
hides the value
Please Note: Activating/deactivating tara clears the max- & min-memories.
1)
Zero-Point Adjustment
:
If there is no pressure or zero-pressure (absolute) applied to the pressure ports the
device will display 0. If there is a permanent deviation (and device is operated
under steady conditions), a permanent zero point adjustment can be carried out. To
carry out the adjustment press button 3 for approx. 5 seconds (Auto Null will be
displayed shortly). The adjustment is done via the OFFSET-value of the sensor
(referring configuration menu).
To recall the manufacturer’s calibration press button 3 for approx. 15 seconds.
Please note: - A zero-point adjustment can only be carried out if the difference between the value on display
is less than 500 digits!
-
If a zero point adjustment was carried out the display shows ”Corr” after a restart .