8-2
8
8.2 GUIDANCE AND MANUFACTURER’S DECLARATION —
ELECTROMAGNETIC IMMUNITY
APPENDIX
The Hausted A.P.C. Series Models and Surgi-Chairs are intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the Hausted A.P.C. Series Models and Surgi-Chairs
should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. if floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
out-put lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(> 95% dip in
U
T
)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(> 95% dip in
U
T
)
for 5 sec
Mains power quality should be that
of a typical commercial or hospi-
tal environment. If the user of the
Hausted A.P.C. Series Models and
Surgi-Chairs require continued
operation during power mains inter-
ruptions, it is recommended that
the Hausted A.P.C. Series Models
and Surgi-Chairs be powered from
an uninterrupted power supply or a
battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A / m
3 A / m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
note
:
U
T
is the a.c. mains voltage prior to application of the test level.