‡ SIGNET 

5900 Salinity Monitor

6.3  WET CAL Procedure

Requirements

Electronic calibration is performed to exacting standards by +GF+ SIGNET.  System calibration will reduce errors which may be caused
by sensor wire lengths longer than the standard 15 foot length.  Wire lengths of 100 feet are acceptable;  cable shield must be
maintained through cable splice.  Calibration may be done by known solution value or by comparison to a refractometer.

Calibration with NIST Traceable Solutions:

When using calibration standards traceable to the National Institute of Standards and Technology (NIST), care must be taken to ensure
the sensor and test solution are at the solution temperature specified on the test solution label.  Care must be taken to prevent
contamination of the calibration solution.  It is recommended to thoroughly rinse the sensor in a small amount of test solution (then
discard) before placing in any test solution for calibration purposes.  The 2-step WET CAL process first allows for verification or
calibration of temperature, followed by verification or calibration of salinity (PPT) using a known process solution.

WET CAL Procedure

 (Solution calibration iIllustrated below)

Note:  Always match test solution manufacturer's temperature recommendation.

Temperature:

+025.0 

°

C >

Temperature:

+025.0 

°

C >

Display:To Change:

To Accept:

Enter known solution salinity or 
reference reading.

Optional: enter zero to reset factory calibration
(zero MUST be re-entered if currently displayed 
to reset factory calibration).

Measured salinity shown 
after temperature calibration.

Press

      to accept salinity

calibration; display returns to 
CALIBRATE MENU in 3 
seconds

Press

       to accept temperature

calibration. 

ENTER

ENTER

quick press

To exit WET CAL at any time without 
saving changes:

Temperature:

+25.7 

°

C >

A) Solution cal.

:  place 

sensor and thermometer into 
standard solution then press 
right arrow key.

A) Solution cal.

:  allow approx. 

3 minutes for stabilization, then 
enter solution temperature.

Salinity
20.0 PPT

Salinity
20.0 PPT

Salinity
20.0 PPT

Содержание SIGNET 5900

Страница 1: ...ered non isolated maximum loop impedance 350 with a 12 V instrument supply voltage 950 with a 24 V instrument supply voltage To isolate output and prevent ground loop problems 1 Use monitor device wit...

Страница 2: ...ents or operating currents that exceed the above specifications 5 2 HIGH alarm mode The relay is active when the solution salinity PPT rises above the setpoint and is inactive when the solution salini...

Страница 3: ...0 100 See dial kit for additional information 6 1 VIEW Menu Menu Displays A G Factory default displays shown in menu column 1 A Active display of salinity PPT and temperature in degrees celsius C or f...

Страница 4: ...int 00 0000 999999 units G Sets relay hysteresis 00 0000 999999 units Set to zero to disable feature H Sets relay minimum and maximum pulse setpoint 00 0000 999999 units I Sets relay pulse rate 000 30...

Страница 5: ...any test solution for calibration purposes The 2 step WET CAL process first allows for verification or calibration of temperature followed by verification or calibration of salinity PPT using a known...

Страница 6: ...A Adjust 3 88 mA Low High ENTER SAVING displayed A B C D E F Contrast 3 Display Decimal Display Average Low Set 4 mA 0 0000 PPT Set 20 mA 40 000 PPT 4 mA Adjust 4 00 mA 4 mA Adjust 4 00 mA Display Dec...

Страница 7: ...acts Max voltage rating 5 A 30 VDC 5 A 125 VAC or 3 A 250 VAC power factor 1 0 Hysteresis User adjustable Current output 4 to 20 mA non isolated internally powered fully adjustable and reversible Upda...

Страница 8: ...40 0000 WET CAL Relay 1 Low 10 0000 PPT Relay 1 Hys 4 0000 Relay 1 Pulse 10 0000 30 0000 Relay 1 Rate 120 Pulses min Last CAL 01 01 98 Sensor type and cell constant Process Units Min Max meter dial r...

Страница 9: ...age 0 5 Off 0 sec Low 4 sec High 8 sec 00 0000 999999 00 0000 999999 3 0 5 0 mA 19 21 mA Celsius or Fahrenheit 0 00 9 99 3 Low 4 sec 00 0000 PPT 40 0000 PPT 4 00 mA 20 00 mA C 2 00 A B C D E F G H I M...

Страница 10: ...fy sensor wiring B Verify instrument temperature input Remove Black and White sensor wires from rear Temp IN and Iso Gnd terminals then place a 1100 resistor across terminals Power instrument and veri...

Страница 11: ...Pg 11 of 12 SIGNET 5900 Salinity Monitor...

Страница 12: ...nue El Monte CA 91731 2882 U S A Tel 626 571 2770 Fax 626 573 2057 For Worldwide Sales and Service visit our website www gfsignet com Or call in the U S 800 854 4090 GEORGE FISCHER Piping Systems 3 59...

Отзывы: