Signet 0250 Setup Tool
Signet 2450 Pressure Sensor Setup
Windows PC
Signet 2450
Pressure
Sensor
USB
GND
Data
+5V out
GND
Data
+5V
2450 Wiring
Operation
1. Select the engineering units for the application: kpa, bar
or psi
2. Set the Pressure measurement where the Loop output
must equal 4 mA.
• Minimum setting 0 psi
• Maximum setting 10, 50 or 250 psi
3. Set the Pressure measurement where the Loop output
must equal 20 mA.
• Minimum setting 0 psi
• Maximum setting 10, 50 or 250 psi
4. Click "Write" to copy the settings from the display into the
Magmeter.
• Click "Save" to save the settings from the display to a
computer fi le.
To use a saved fi le:
• Click "Load"
• Navigate to the saved fi le
• Select Open
• Click "Write"
• Click "Read" to confi rm
Signet 2750 pH Sensor Setup
2750 Wiring
2750 Operation
1. Select the engineering units for the application: pH or ORP
2. Set the pH or ORP measurement where the Loop output
must equal 4 mA.
• Minimum setting 0 pH, -1000 mV ORP
• Maximum setting 14 pH, 2000 mV ORP
3. Set the pH or ORP measurement where the Loop output
must equal 20 mA.
• Minimum setting 0 pH, -1000 mV ORP
• Maximum setting 14 pH, 2000 mV ORP
4. Click "Write" to copy the settings from the display into the
Magmeter.
• Press "Save" to save the settings from the display to a fi le.
To use a saved fi le:
• Click "Load"
• Navigate to the saved fi le
• Select Open
• Click "Write"
• Click "Read" to confi rm.
Windows PC
Signet 2750
pH/ORP
Sensor,
Submersible
+5V in
Data
GND
1
2
3
4
+
S
-
Signet 2750 pH/ORP
Sensor, Inline
USB
GND
Data
+5V out
GND
Data
+5V out
GND
Data
+5V