Signet 9900 Modbus Module

Register

Size Read/Write Data Type Description

49001

1

Read/Write

UINT

Modbus Address 1 to 247 (Default 34)

49002

1

Read/Write

UINT

Network Termination, 1 = ON 0 = Off  (Default 0)

49003

1

Read/Write

UINT

Modbus Mode 0 Modbus RTU 1 Modbus ASCII (Default 0)

49004

1

Read/Write

UINT

Baud Rate 0 = 1200, 1 = 2400, 2 = 4800, 3 = 9600, 4 = 19200, 
5 = 38400, 6 = 57600, 7 = 115200 (Default 4) 

49005

1

Read/Write

UINT

Parity 0 = No Parity, 1 = Odd Parity, 2 = Even Parity (Default 2)

49006

1

Read/Write

UNIT

Floating Point Little Endian = 0, Big Endian = 1 (Default 1)

49007

1

Read/Write

UINT

Write 0 (zero) to save registers, Write Modbus Address to save 
registers and reboot

Register

Size Read/Write Data Type Description

40001

2

Read

UINT

Status

40003

2

Read

Float

Primary Reading

40005

2

Read

Float

Secondary Reading

40007

1

Read

UINT

Measurement Type

40008

1

Read

UINT

Primary Units of Measure Code

40009

1

Read

UINT

Secondary Units of Measure Code

Register

Size Read/Write

Data Type Description

49104

1

Read

UINT

9900 HSB Transactions 

49105

1

Read

UINT

Second Counter

49106

1

Read

UINT

Bus Message Counter

49107

1

Read

UINT

Bus Message Error Counter

49108

1

Read

UINT

Slave Exception Counter

49109

1

Read

UINT

Slave Message Counter

49110

1

Read

UINT

Slave No Response Counter

49111

1

Read

UINT

Slave NAK Counter

49112

1

Read

UINT

Slave Busy Counter

49113

1

Read

UINT

Bus Character Overrun Counter

Live Reading

Communication Settings

Communication Diagnostic Registers

49501

15

Read

STR

Vendor Name “Georg Fischer Signet”

49516

10

Read

STR

Product Code “3-9900-1x”

49526

10

Read

STR

Major/Minor Revision 9900 Transmitter “xx-xxX”

49536

15

Read

STR

Vendor URL “www.gfps.com”

49551

10

Read

STR

Product Name “SmartPro”

49561

10

Read

STR

Model Name “9900”

49571

10

Read/Write

STR

User Application Name

49581

5

Read/Write

STR

Tag Name

49586

5

Read/Write

STR

Location

49591

5

Read

STR

9900 Serial Number “1204240438” Leading “6” dropped

49596

5

Read

STR

Modbus Module Serial Number “111317”

49601

1

Read

UINT

Manufacture Code 0x0032

49602

2

Read

UINT

Module Version 0x30010004  Version 1.00.04

Device Identifi cation

Modbus Register Map

Содержание 3-9900.270-MX

Страница 1: ...efore or after the base unit is mounted If the 9900 Base Unit will be mounted using the accessory wall mount kit 3 9900 392 install plug in modules rst If the Direct Conductivity Resistivity Module or...

Страница 2: ...y Information Product Registration Safety Information CAUTION Exercise care when installing module Do not bend connecting pins Align pins and connectors then push module firmly into place Avoid Electr...

Страница 3: ...White Gray D0 RS485 Pin 2 D1 RS485 Pin 1 V 12 to 24 VDC Pin 3 V 12 to 24 VDC Common Pin 4 Com Reset GND Pin 5 Brown D0 RS485 Yellow D1 RS485 Black V 12 to 24 VDC White V 12 to 24 VDC Common Gray Com...

Страница 4: ...te Data Type Description 49104 1 Read UINT 9900 HSB Transactions 49105 1 Read UINT Second Counter 49106 1 Read UINT Bus Message Counter 49107 1 Read UINT Bus Message Error Counter 49108 1 Read UINT Sl...

Страница 5: ...gister 49005 0 No Parity 1 Odd Parity 2 Even Parity Default 2 Set baud Rate Register 49004 0 1200 1 2400 2 4800 3 9600 4 19200 5 38400 6 57600 7 115200 Default 4 Write Modbus Address to Register 49007...

Страница 6: ...5102 pH Temperature ORP 0x5103 ORP Raw mV Cond Res 0x5704 Cond Res Temperature Pressure 0x4105 Pressure NA Level Volume 0x5C06 Level Volume Temperature 0x4007 Temperature NA 4 to 20 mA Input 0x0008 Sc...

Страница 7: ...e BPH 134 242 Barrels per Hour BPD 135 Barrels per Day FT S 20 Feet per Second M S 21 Meters per Second All Others 252 All other ow units Flow Totalization Secondary All Units 240 pH Primary pH 59 pH...

Страница 8: ...ssembly for M2 3 9900 270 CB3 159 200 125 Replacement M12 Connector Assembly for M3 3 9900 270 CB4 159 200 129 Replacement Cable Assembly for M4 5541 5005 159 855 021 5 meter 16 ft M12 cable 5541 5010...

Отзывы: