001-14154-33-EN rev. 2
6 of 8
2.5 Adjust the Charger Detection Time Limit
NOTE
: Not available for Maxi Sky 600 ECS models
Adjusting the charger detection time limit will restrict the time the unit is
allowed to detect the charge station when the Return to Charge button is
pressed, stopping the motor after the selected time delay.
A) Enter into the Technical Service Mode and select the fifth choice (consult section 2).
B) Use the LEFT button to select the charger detection time limit during the return to charge as per the
TIME ADJUSTMENT chart below. Each push of the button allows you to go from one section to the
next:
LIFT MODELS
TIME ADJUSTMENT
All Models
1 beep*
2 beeps
3 beeps
4 beeps
2.5 min
5 min
7.5 min
10 min
* Default setting from factory
C) To confirm the time adjustment of the charger detection, press on the PROG button.
D) Now you can choose a new selection by pressing on the UP button on the hand control. To return to
normal use, turn off the lift and turn it back on.
2.6 Reading of the Service Counter
The Service Counter allows you to know the total number of lifts the unit has
performed since its manufacturing or resetting. One cycle is recorded for
every approximately 60 cm (24 in) the strap travels in the up direction, then
approximately 60 cm (24 in) it travels in the down direction.
2.6.1 For units not equipped with a plastic insert switch and all Maxi Sky 1000 / V10
A) Enter the Service Technical Mode and select the sixth choice (consult section 2).
B) The number of beeps will equal to the cumulative lifts since the last reset of the service counter.
Each beep equals to 250 cycles (see chart below)
.
Beeps
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Cycles
250
500
750
1000
1250
1500
1750
2000
2250
2500
2750
3000
3250
3500
3750
Beeps
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
Cycles
4000
4250
4500
4750
5000
5250
5500
5750
6000
6250
6500
6750
7000
7250
7500
The unit begins to
produce audible
warning beeps.
Lift service
indicator is
turned on