Maintenance
How often maintenance will be required depends to a great extent on the quality
of the water supply and how often the machine is used. The period between
each maintenance should be determined in each particular case. We recommend,
however, that the following maintenance measures should be performed once or
twice every year.
Periodic maintenance
May only be performed by authorized personnel.
This machine is connected to mains power supply and some
components are live.
Component
Interval
Inspect
Replace as
Replace
Replace every
every year
required
every year
two years
Door gasket
•
•
Hoses between
•
•
dosing pump
and detergent bottle
Hose between dosing
•
•
pump and machine
Hose in hose pump
•
•
Filter in input media
•
•
Overheating protection
•
•
Spray arm bearings
•
•
Sterile filter in dryer
•
•
•
Thermostat in dryer
•
•
Hoses to dryer
•
•
Printer ribbon
•
•
option