4.
VALIDATION MONITOR, PRINTER (OPTION)
(OPTION ONLY FOR 1400-1600LXP/1800LXA MODELS)
The validation monitor has its own probes to check and validate every parameter of the cycle of the washing cycle.
The validation monitor option includes:
One built-in panel printer
One pressure transmitter for verification of proper functioning of the recirculating pump
Two flow-meters measure the exact quantity of chemical taken by the machine during the cycle
One water level pressure stat for verification of proper filling of the sump
Two dual temperature probes for verification of proper temperature values
One electrical link to the heating system for verification of proper temperature inside the chamber
One electrical link to the draining system for verification of proper water exhaust between two wash cycle phases
EXAMPLE OF TICKET