Geotech Biogas Скачать руководство пользователя страница 11

Biogas  Operating Manual 

Page 10 

Geotechnical Instruments (UK) Ltd 

BMUK0001 / Issue 1 

You can do this by pressing key 

to re-set the reading procedure or by pressing 

store reading. 

Either option allows a new reading sequence to begin. 

2.  At this point connect the sample tube (with water trap) from the sample point to the inlet port of the 

instrument, ensuring the connector ‘clicks’ into place. If using a gas pod, it should plugged-in to 
external port 1 and connected to the outlet of the instrument. At this point you will see the system 
pressure reading change as it responds to the pressure on the inlet hose. It is important to 
remember that the system pressure reading is only active during this initial phase. After the pump 
is activated or if the user presses 

measure flow or 

store reading the system pressure is 

frozen until the next reading cycle has begun.  

3. Pressing 

the  pump key draws a gas sample into the instrument. You should see the main gas 

readings start to change. It is recommended to run the pump until the gas readings have stabilised 
(typically 30 seconds) using the counter provided. If you are only interested in the main gases 
pressing the 

store key now would record these values and re-start the reading cycle. Note…If 

you have enabled gas alarms a flashing bell symbol may be displayed next to the appropriate gas 
reading value if a pre-set alarm condition has been passed. 

4.  Assuming you have selected a flow device (via instrument settings) press key 

to take a flow 

reading. At this point you should see a brief message confirming that the gas readings have been 
fixed (i.e. stored temporarily). Depending on which flow device you have selected the operation 
may vary. For example if you have no flow device selected then you will not have a key 

option, 

pressing the 

store key will store the currently displayed data to memory.   

5.  If enabled (via instrument settings) you will be prompted to enter a temperature reading or attach a 

temperature probe. If you are not using an external gas pod you might want to disable this option. 
You can then plug a temperature probe in to external port 1 (in place of the gas pod) at the start of 
the reading cycle. The temperature reading is then displayed on the main gas screen and will be 
frozen when key 

is pressed to start the flow measurement phase. Otherwise simply follow the 

on-screen instructions to record a valid temperature reading either by attaching a probe or 
manually entering the temperature recorded using a third party sensor. If no temperature is 
available then 20 degrees centigrade is assumed for the subsequent flow calculations.     

6.  If you have selected either pitot tube or orifice plate as your device type you will now be asked to 

zero the static and differential pressure transducers. It is possible to skip this stage but it is 
important to zero the transducers for repeatable results especially on very low flows. To zero the 
transducers simply follow the on-screen instructions. Remember you will have to disconnect the 
hoses from the sample point and wait a few seconds for the readings to stabilise before zeroing. 
You will then be returned to the main read screen where the flow calculation will be displayed along 
with the current static and differential pressures. You will also be able to review the Orifice and 
pipe diameter that are currently selected. These can be changed (via instrument settings) if 
required. Once the flow reading has stabilised simply press store 

to record the reading.  

7.  If you have selected an anemometer as your device type there is no need to zero the transducers 

and this stage is skipped automatically. You be returned directly to the main read screen where the 
velocity and flow will be displayed along with the currently selected pipe diameter (required for a 
valid flow calculation). The anemometer should be plugged into external port 2 on the instrument. 
After it has been correctly inserted into the sample point, allow the flow reading to stabilise and 
press store 

to record the reading. 

8.  If you have selected user input as your device type you will be prompted to enter a flow reading 

manually. After typing in the flow reading you can return to the main reading screen and the flow 
measurement you entered will be displayed. Press store 

to complete the reading. 

9. After 

the  store key is pressed and if enabled (via instrument settings) you will be prompted to 

enter an 8 character identification code for the current reading set. This is done by using the  
cursor keys to navigate around a grid of alphanumeric characters. The 

enter key is used to 

select each character as required. Key 

is used complete the process and exit this screen. At 

anytime the 

backspace key can be used to erase any selected characters if required. 

Содержание Biogas

Страница 1: ...Biogas Operating Manual Geotechnical Instruments UK Ltd Sovereign House Queensway Leamington Spa Warwickshire...

Страница 2: ...EADING MEMORY 8 3 11 VIEW READINGS 8 3 12 ADJUST CONTRAST 8 3 13 INSTRUMENT SETTINGS 8 3 14 SET TIME DATE 8 3 15 SET GAS ALARMS 8 3 16 OPERATING LANGUAGE 9 4 TAKING READINGS 9 4 1 PRELIMINARY CHECKS 9...

Страница 3: ...ESETZT WERDEN DARF 29 10 8 KALIBRIERUNG 29 10 9 WARTUNG 29 10 10 EG KONFORMIT TSERKL RUNG 30 11 INSTRUCTIONS FOR SAFE USE FRENCH LANGUAGE 31 11 1 INSTRUCTIONS SPECIFIQUES POUR LES INSTALLATIONS DANS D...

Страница 4: ...e precautions e g regular checks as part of routine inspections or establishing from the material s data sheet that it is resistant to specific chemicals that prevent it from being adversely affected...

Страница 5: ...uld be treated as such 2 1 Main features of the Instrument Measurement of the following gases Methane and Carbon Dioxide by Infra Red Absorption Oxygen by chemical cell Other gases possible by way of...

Страница 6: ...mined self test sequence taking approximately 30 40 seconds During this time many of the analyser s functions are tested including General operation Pump function Gas flow measurement Calibration Back...

Страница 7: ...lculation Where there is no data is available dashes are displayed This usually occurs when a particular reading or parameter has been skipped by the user 3 5 Cold Start Note This function should only...

Страница 8: ...g Screen The main reading screen is also considered to be the normal operation screen and most operations are carried out from this starting point All of the instrument readings are displayed in this...

Страница 9: ...screen allows the user to configure the reading sequence and various other options as desired When selected the current settings are displayed These include the following parameters Purge Time Prompt...

Страница 10: ...ith a known concentration calibration check gas DO s Travel to site with the instrument in the vehicle s interior not in the boot where it may be subjected to extremes of temperature and possible shoc...

Страница 11: ...start the flow measurement phase Otherwise simply follow the on screen instructions to record a valid temperature reading either by attaching a probe or manually entering the temperature recorded usin...

Страница 12: ...be Reading The Biogas instrument has the facility to automatically display and record the borehole temperature via an optional temperature probe When a temperature probe is fitted to external port 1 R...

Страница 13: ...igher never lower than the actual methane concentration being monitored The extent to which the methane reading is affected depends upon the concentration of the methane in the sample and the concentr...

Страница 14: ...r is calibrated when a known quantity of the target gas is present 5 2 Calibration Gases User calibration of a Gas Analyser will improve the data accuracy in the range of the calibration gases used Th...

Страница 15: ...of the exhaust port of the analyser or in cases of over pressurisation the 1 16 port on the pressure relief valve It is recommended that both ports have exhaust tubing attached The exhaust tubing mus...

Страница 16: ...tive and will be shown as N A Step 1 Enter Target Concentrations DO NOT CONNECT CALIBRATION GAS YET Enter the certified concentrations of your calibration gases For two gas mixtures that contain 0 0 O...

Страница 17: ...libration equipment as pictured above and turn on gas at regulator Do not run the pump Allow the gas to flow through the unit for approximately two minutes or until reading stabilises Press to select...

Страница 18: ...ing process If it will not zero then refer to the instructions given in the Factory Settings section If the Gas Analyser continues to fail in zeroing then the unit must be returned to Geotechnical Ins...

Страница 19: ...Last Field Calibration This option can be found in the calibration menu of the field calibration section When selected this option will display the time and date that the last field calibration was p...

Страница 20: ...plied with your instrument The following parts are supplied by Geotechnical Instruments and can be user serviced In line water filter This should be regularly inspected for obstructions or damage and...

Страница 21: ...pproximately 3 years in air Range CH4 0 70 to specification 0 100 reading CO2 0 40 to specification 0 100 reading O2 0 25 Typical accuracy Gas 0 5 vol 5 15 vol 15 FS CH4 0 5 1 0 3 0 CO2 0 5 1 0 3 0 O2...

Страница 22: ...om calibration pressure Barometric pressure accuracy 5mbar typically PHYSICAL Weight 2 Kilograms Size L 63mm W 190mm D 252mm Case material ABS Keys Membrane panel Display Liquid crystal display 40 x 1...

Страница 23: ...6 1996 EN 50270 1999 Electromagnetic compatibility EMC Generic standards Emission standard for industrial environments Electromagnetic compatibility EMC Testing and measurement techniques Radiated ra...

Страница 24: ...ai dispositivi che sono contrassegnati con il numero di certificati Sira 06ATEX2202X Se si prevede che l apparecchiatura venga a contatto con sostanze aggressive ad esempio liquidi o gas acidi che pot...

Страница 25: ...guasto Se una taratura errata dell utente ha determinato la visualizzazione di un avvertenza il problema potrebbe risolversi ripristinando lo strumento sulle impostazioni di fabbrica impostandolo su z...

Страница 26: ...gode della certificazione ATEX per l uso nelle aree potenzialmente esplosive Di conseguenza va sottoposto a manutenzione esclusivamente da parte di tecnici qualificati Il mancato rispetto di questa in...

Страница 27: ...QRC 9 10 Dichiarazione di conformit EC Prodotti Biogas Analizzatore gas Landfill Geotechnical Instruments UK Ltd dichiara che gli elementi sopra descritti sono conformi ai seguenti standard Direttiva...

Страница 28: ...mit der Zertifikatsnummer Sira 06ATEX2202X versehen sind Besteht die Wahrscheinlichkeit dass das Ger t in Kontakt mit aggressiven Stoffen gelangt z B saure Fl ssigkeiten oder Gase die Metalle angreife...

Страница 29: ...NING vorangestellt dem eine relevante Beschreibung folgt Der wahrscheinlichste Grund f r die Fehlermeldungen ist entweder eine falsche Benutzerkalibrierung oder ein Sensordefekt Falls eine falsche Ben...

Страница 30: ...atz potenziell gef hrlicher explosiver oder giftiger Gase m ssen alle notwendigen Sicherheitsvorkehrungen getroffen werden Vor der Durchf hrung der Arbeiten muss f r jedes eingesetzte Gas das entsprec...

Страница 31: ...dass Messschl uche nicht verschmutzt oder besch digt sind QRC Anschl sse berpr fen Sie regelm ig dass die O Ringe an den QRC Gasanschl ssen nicht besch digt sind Ein besch digter O Ring kann Luft in d...

Страница 32: ...ccord A qu des appareils portant la marque de num ro de certificat Sira 06ATEX2202X Si le mat riel est amen tre en contact avec des substances agressives par exemple des liquides ou des gaz acides sus...

Страница 33: ...utilisateur ou d une d faillance des capteurs Un talonnage incorrect peut tre rectifi en remettant l appareil sur les valeurs r gl es en usine en effectuant une remise z ro ou l talonnage requis pour...

Страница 34: ...non respect de cette prescription entra nera l annulation de la garantie et risque d annuler la certification ATEX Avertissement Si l analyseur fait l objet d un entretien par des techniciens non qual...

Страница 35: ...s aux normes suivantes Directive ATEX 94 9 CE Organisme de certification SIRA Certification Service Service de certification SIRA Num ro d organisme notifi 0518 Adresse Rake Lane Eccleston Chester CH4...

Страница 36: ...l conector A a dispositivos marcados con el certificado Sira N 06ATEX2202X Si existe la posibilidad de que el equipo entre en contacto con sustancias agresivas como por ejemplo gases o l quidos cidos...

Страница 37: ...o del sensor Si la calibraci n de usuario es incorrecta y ha provocado la advertencia debe corregirse corrigiendo el instrumento a los ajustes de f brica poni ndolo a cero o efectuando una calibraci n...

Страница 38: ...iento de esta norma anular la garant a y podr a invalidar la certificaci n ATEX Advertencia Si el analizador es revisado por t cnicos no cualificados la certificaci n ATEX podr a quedar invalidada y e...

Страница 39: ...uientes normas Directiva ATEX 94 9 EC Organismo de Certificaci n Servicio de Certificaci n SIRA N del Organismo Notificado 0518 Direcci n Rake Lane Eccleston Chester CH4 9JN N de Certificado Sira 06AT...

Страница 40: ...ment directive WEEE makes producers responsible from July 1 st 2007 in meeting their obligations with the fundamental aim of reducing the environmental impact of electrical and electronic equipment at...

Страница 41: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: