2.3 Sensor Installation with tool
The installation tool (3-2550.355) simplifies
installation and insures proper insertion depth for
maximum accuracy.
3
The 2550 Magmeter weighs
11 lbs. Care must be taken
to properly support the unit
with straps and/or bracing
when necessary.
1. Set installation tool angle
plate (1A) on pipe next to fitting.
Insert adjusting slide (1B) into
top of angle plate.
Installation Tool
Front View
1A
1B
FLOW
PIPE I.D.
2A
Process Pipe
Cutaway Side View
2B
2. Place height gauge (2A) short end inside pipe
and pull up against inside diameter of pipe. Insert
height gauge long end into slot in adjusting slide.
Tighten wing nut (2B) to hold adjusting slide in
place against angle plate. Remove height gauge
from pipe.
3. Loosen the three bolts at
top of 2550 adjustment nut
(3A). Wrap the adjustment
nut threads with three full
wraps of sealing tape and
insert sensor into fitting until
base of the electronics
enclosure is resting on top
of adjusting slide (3B). Be
sure cable ports are point-
ing downstream. Thread
adjustment nut securely into
pipe. Tighten all three bolts
to secure adjustment nut.
Important Safety
Information!
3A
Process Pipe
Cutaway Front View
3B
Adjustment
Nut
Do not remove safety
ring or adjustment nut
from sensor.