3.3.2 Interior diameter
You must enter the precise interior diameter of your
pipe as measured during the installation process.
Entry is made in inches when the dimension
entered is less than 40, and automatically switches
to millimeters when the dimension entered is
greater than 40.
To change the id selection, enter the access code
(section 3.2.1) then press the down arrow until the
display shows:
Arrow
keys advance main menu function;
run/edit
key returns to normal operation.
12
ENTER
ENTER
ENTER
ENTER
ENTER
3.
2.
1.
4.
7.
6.
5.
8.
9.
10.
11.
12.
ENTER
13.
The example illustrates setting
id to 10.40 in.
The decimal point location
is selected after the last
flashing digit is set.
Down arrow key
changes flashing digit.
The factory default setting
for id is 2.067 in.
ENTER key stores new setting
and advances flashing digit.
Down arrow key
changes flashing digit.
Up arrow key changes
flashing digit.
ENTER key stores final setting
and returns to main menu.
Down arrow key moves
decimal to right.
Step:
Press:
To display: