Zoom80
| 100
Check & Adjust
The following table explains the most common settings.
Step
Description
1.
Place and secure the instrument into the tribrach and onto a tripod.
2.
Using the tribrach footscrews, level the instrument with the elec-
tronic level. SHIFT (F12) to access STATUS Level & Laser
Plummet.
3.
PAGE
(F6) to access the Laser Plummet page. Switch on the laser
plummet. Inspection of the laser plummet should be carried out on a
bright, smooth and horizontal surface, like a sheet of paper.
4.
Mark the centre of the red dot on the ground.
5.
Turn the instrument through 360° slowly, carefully observing the
movement of the red laser dot.
The maximum diameter of the circular movement described by the
centre of the laser point should not exceed 3 mm at a distance of
1.5 m.