Zoom80
| 96
Check & Adjust
4.5 Adjusting the Circular Level of the Instrument and
Tribrach
Adjusting the
circular level
step-by-step
Step
Description
1.
Place and secure the instrument into the tribrach and onto a tripod.
2.
Using the tribrach footscrews, level the instrument with the elec-
tronic level. SHIFT (F12) to access STATUS Level & Laser
Plummet.
3.
Check the position of the circular level on the instrument and
tribrach.
2
1
4
4
F12
SHIFT