
10
5. TROUBLESHOOTING
Maintenance and troubleshooting of pressure cells is confined to periodic checks of cable
connections and maintenance of terminals. Once installed, these instruments are usually
inaccessible and remedial action is limited.
Gauges should not be opened in the field.
Should
difficulties arise, consult the following list of problems and possible solutions. Return any faulty
gauges to the factory. For additional troubleshooting and support contact Geokon.
Symptom: Pressure Cell Readings are Unstable
Is the readout box position set correctly? If using a datalogger to record readings
automatically are the swept frequency excitation settings correct?
Is there a source of electrical noise nearby? Likely candidates are generators, motors, arc
welding equipment, high voltage lines, etc. If possible, move the instrument cable away from
power lines and electrical equipment or install electronic filtering.
Make sure the shield drain wire is connected to ground whether using a portable readout or
datalogger. Connect the shield drain wire to the readout using the blue clip. (Green for the
GK-401.)
Does the readout work with another pressure cell? If not, it may have a low battery or
possibly be malfunctioning.
Symptom: Pressure Cell Fails to Read
Is the cable cut or crushed? Check the resistance of the cable by connecting an ohmmeter to
the gauge leads. Table 1 shows the expected resistance for the various wire combinations.
Table 2 is provided for the user to enter the actual resistance found.
Cable resistance is approximately 14.74
Ω
per 1000' of 22 AWG wire. If the resistance is very
high or infinite (megohms) the cable is probably broken or cut. If the resistance is very low (
<
20
Ω
) the gauge conductors may be shorted. If a cut or a short is located in the cable, splice
according to the instructions in Section 2.3.
Does the readout or datalogger work with another gauge? If not, it may have a low battery or
possibly be malfunctioning.
Symptom: Thermistor resistance is too high
Likely there is an open circuit. Check all connections, terminals, and plugs. If a cut is located
in the cable, splice according to instructions in Section 2.3.
Symptom: Thermistor resistance is too low
Likely there is a short. Check all connections, terminals, and plugs. If a short is located in the
cable, splice according to instructions in Section 2.3.
Water may have penetrated the interior of the transducer. There is no remedial action.