7
EN
Technical Data
Measuring range timber
6 - 44 %
Accuracy timber
± 1 %*
Measuring range construction material
0,2 - 2 %
Accuracy construction material
± 0,05 %
Power supply
3 x Cr2032 batteries
Dimensions
130 x 40 x 25 mm
Weight
102 g
C
FEATURES
•
Measuring principle electrical resistance
•
Length of electrodes 8mm Impact-proof housing
•
Automatic shut-off
•
Protection cover with integrated function and battery check
•
Replaceable electrodes
D
FUNCTION CHECK
Connect electrodes with T contacts of the protective cap: „O-----T-----O“.
Display timber
27 % ±2 %
Display construction material 1,25 % ±0,1 %
If those values are not displayed instrument should be sent for service.
Please handle measuring instruments with care.
Clean with soft cloth only after any use. If necessary damp cloth with some water.
If instrument is wet clean and dry it carefully. Pack it up only if it is perfectly dry.
Transport in original container / case only.
Change of electrodes:
Unscrew all 5 screws on the back of the instrument and remove housing lid. Unscrew all 5 screws on the
electronic plate and replace the electrodes. Close instrument again.
E
CARE AND CLEANING
*adjusted to european timber at 20°C