S-
LINK User’s Manual Revision 2.8
53
8
Appendix D
8.1 S-LINK-2 BINARY MODE DESCRIPTION
8.1.1 DESCRIPTION
The S-LINK-2 has two Communication Modes: the Binary Mode for FAST data acquire and the ASCII
mode for slow and high resolution values. The resolution of the Wattmeter Mode is 11- bit, the resolution
of the Joulemeter Mode is 12-bit. The resolution in ASCII mode is 15 bits.
The value in binary mode is coded in two bytes
JOULEMETER MODE
Byte #1
C
O
2
X
X
X
X
X
Byte #2
C
O
X
X
X
X
X
X
2 = The MSB of the 12 bit codification for the Joulemeter Mode.
X = Bits of the Codification.
WATTMETER MODE
Byte #1
C
O
2
Y
X
X
X
X
Byte #2
C
O
X
X
X
X
X
X
2 = The sign.
Y = The MSB for the Wattmeter Mode.
X = Bits of the Codification.
FOR ALL MODES
C = Channel number
0 = Channel #1
1 = Channel #2
O = Order
1 = First byte
0 = Second Byte
Содержание S-LINK
Страница 1: ......
Страница 25: ...S LINK User s Manual Revision 2 8 20 Go to USB port properties Select Port parameters and select Advanced ...
Страница 38: ...S LINK User s Manual Revision 2 8 33 Select a new Com Port and click ok ...
Страница 41: ...S LINK User s Manual Revision 2 8 36 Go to Properties ...
Страница 42: ...S LINK User s Manual Revision 2 8 37 Go to TCP IP Properties ...
Страница 46: ...S LINK User s Manual Revision 2 8 41 ...
Страница 48: ...S LINK User s Manual Revision 2 8 43 Select Server option set the CPU performance to high and click OK ...
Страница 60: ...S LINK User s Manual Revision 2 8 55 ...