Troubleshooting.
!TRIG Message: The repetition rate of your laser is too high for you PC performance, or you are triggering
on noise. Set the trigger to a higher level or upgrade your PC.
The Application is not responding when I select Status or when I want to change the Graphic settings:
Press the ALT key, your status window may be behind the PCS-Link application. This happens when you
set the time stamp of the graphic to a high value on certain PCs.
Cannot Connect to the Ethernet Port.
1_) Reboot the computer when you change the IP address if you have a direct connection. 2)
One possible cause is the Firewall on this machine is blocking this port. Press the 'Add Rx Port' button to
add this port to the Firewall. If the button caption reads 'Remove Rx Port' then the port has already been
added and can be removed by pressing this button.
3) Disable your local Firewall.
4) Read the Lantronix FAQ.
http://ltxfaq.custhelp.com/cgi-bin/ltxfaq.cfg/php/enduser/std_alp.php