![Gentec 203605 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/gentec/203605/203605_user-manual_2213625049.webp)
Beamage-M² User Manual
Revision 6
42
7.10.
OTHER ISSUES
NOTE: The Beamage must not be disconnected while streaming.
1. Curve fit passes below 0
a. The measurement is made too far away from the waist:
i. We have too many points beyond -3.5Zr and 3.5Zr.
ii. We do not have enough points between
–Zr and Zr.
iii. Solutions:
1. Remove points far from the waist, one by one, and then recalculate.
2. Note the values of Z0 and Zr, clear all results and remeasure a set of
points expanding ONLY from -3.5Zr to +3.5Zr, not beyond.
3. Take more points around the waist.
b. The software cannot find the waist:
1. See section 7.8 Lens choosing guide above.
2. The value of M² is below 1
a. The system (not laser) affects the beam:
i. Check and clean every mirror and ND filter.
ii. Dust can change drastically a M² measurement.
iii. Broken optics or mirrors can change drastically a M² measurement.
iv. Look and ask for collimator adjustment.
b. ISO minimum requirements are not respected:
i. Look at the M² display for ISO minimum requirements.
ii. If not respected, add more data points (MORE THAN REQUIRED BY ISO).
c. The measurements were made too far away from the waist:
i. See section 7.8 Lens choosing guide above.
3. The software continually asks to change attenuation and take no measurement
a.
Make sure the laser beam illuminates the camera’s sensor.
b. The laser beam might be so intense that it saturates the sensor near the focus.
Possible solutions:
i. Change the lens for a longer focal length, this will widen w
0
.
ii. Add attenuation before M1; a beam that has a power beyond 100 mW might
have some saturation issues near the focus.
c. The computer might be too slow (very low frame rate):
i. See section 7.5 Tips to increase the frame rate.
ii. Use the
Image averaging
function to stop the flickering.
4. Other things to double-check:
a. The mirrors are well set.
b. The camera is securely fastened.
c. The filter on the camera is completely screwed.
d. The moving stage is plugged with a
special
USB cable.
e. The beam is sufficiently attenuated.
f.
The drivers and software are up to date.
g. In case of uncertainty, uninstall everything with Windows control panel and install the
latest versions.
Содержание 203605
Страница 1: ...Beamage M User Manual Revision 6...
Страница 58: ...Beamage M User Manual Revision 6 51...