![Gentec 203605 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/gentec/203605/203605_user-manual_2213625035.webp)
Beamage-M² User Manual
Revision 6
28
If the quadratic fit curve passes under zero, the PC-Beamage software will try to solve the problem by
deleting the measurements farthest from
𝑍0
and by automatically recalculating the value of M². The PC-
Beamage software will also offer a recommended travel distance between
−3.5𝑧𝑅
and
3.5𝑧𝑅
to correct the
problem (see Figure 5-25).
Figure 5-25:
Curve Fit Passes Under Zero.
A minimum of 5 beams must be detected to calculate M
2
.
Upon clicking the
START ISO SCAN
button, click
Yes
if you want to save previous data or
No
to ignore
them. The data erase themselves and then a new M² measurement begins automatically.
Figure 5-26:
Save previous data
Once the new measurement is completed, click the
Calculate M²
button to display the
M² Results
tab.