Gentec-E PE10B-Si Скачать руководство пользователя страница 15

 

Version 2.0

 

 

 

 

 

PH Series Instruction Manual 

 

Gentec Electro-Optics Inc.  All rights reserved

 

12 

Slope: The slope of the linear interpolation. 

Sens

LOW_

λ

: Sensitivity at 

LOW

Sens

HIGH_

λ

: Sensitivity at 

HIGH

LOW

: The next lowest available wavelength near your desired wavelength. 

HIGH

: The next highest available wavelength near your desired wavelength. 

DESIRED

: Desired Wavelength. 

:  The difference between the desired wavelength and the inferior wavelength. 

 

 

4.5.1  Example

 

 
You have a PH100-Si and your laser is at 632.8 nm.  
 
See your CERTIFICATE OF CALIBRATION for the sensitivity of your power detector as a function of 
the wavelength.   
 

Wavelength (nm) 

Sensitivity (A/W) 

620 

0.32 

630 

0.35 

640 

0.37 

650 

0.40 

660 

0.43 

   
632.8 nm is between 630 nm and 640 nm therefore, 
 

Sens

LOW_

λ

 = 0.35 A/W 

Sens

HIGH_

λ

 = 0.37 A/W 

LOW

 = 630 nm 

HIGH 

= 640 nm 

:  632.8 - 630 = 2.8 nm 

Slope: (0.37-0.35)/(640-630) = 0.002 

Sensitivity 

desired_

:  0.35 + 2.8 * 0.002= 0.356 A/W 

 
 

Содержание PE10B-Si

Страница 1: ...to obtain a return authorization number Return the material to the appropriate address below Contacting Gentec Electro Optics Inc To help us answer your calls more efficiently please have the model n...

Страница 2: ...idity or surface contaminants exposure to flame solvents or water and connection to improper electrical voltage TABLE OF CONTENTS WARRANTY I First Year Warranty i Lifetime Warranty i TABLE OF CONTENTS...

Страница 3: ...TO DETECTOR DIMENSIONS 3 FIG 1 2 TYPICAL GE TEMPERATURE DEPENDENCE VS THE WAVELENGTH 10 FIG 1 3 TYPICAL SI AND SIUV TEMPERATURE DEPENDENCE VS THE WAVELENGTH 10 FIG 1 4 PH100 SI TYPICAL SPECTRAL RESPON...

Страница 4: ...uoted maximum ranges are average power at the 1064 nm wavelength for PH100 Si and PH20 Ge 532 nm for PH100 SiUV and 850 nm for PH100 SiUV with OD 2 quoted minimum ranges are average power at the 980nm...

Страница 5: ...ific PH and PE Series head Wavelengths are not available where the sensor physical properties don t allow it The Gentec eo monitors use the data in this connector to adjust their characteristics autom...

Страница 6: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 3 1 2 2 Dimensions FIG 1 1 PHOTO DETECTOR DIMENSIONS...

Страница 7: ...30 mW 320 mW cm2 1064 nm Minimum Powera 0 6 nW 980 nm 0 6 nW 850 nm 2 nW 1550 nm Damage threshold Maximum Average Power Density 100 W cm 2 100 W cm 2 100 W cm 2 Typical Detector saturation currentb 6...

Страница 8: ...W 980 nm 0 45 A W 850 nm 0 98 A W 1550 nm With attenuator PH100 Si OD1 OD2 PH100 SiUV OD1 OD2 PH20 Ge OD1 OD2 Maximum Power typ w OD 1 300 mW 1064 nm 25 mW 532 nm 300 mW 1064 nm Minimum Power a w OD 1...

Страница 9: ...measurable power M Link 200 W S Link 175 W Maestro 150 W 633nm M Link 32 W S Link 26 W Maestro 22 W 1047 nm Maximum Average Power Density 65 mW cm2 532 nm 320 mW cm2 1064 nm Noise equivalent power 10...

Страница 10: ...Repetition Rate 1000 Hz Max Pulse Width 10 s Maximum Measurable Energy M Link 0 16 J S Link 0 13 J Maestro 0 11 J 633nm M Link 3 2 nJ S Link 2 6 nJ Maestro 2 4 nJ 1047 nm Maximum Energy Density 5 J c...

Страница 11: ...ack protective cover on the detector head before making the offset nulling If you want to subtract the room light do the offset nulling with the cover off 2 2 QUICK POWER MEASUREMENT PROCEDURE This se...

Страница 12: ...component not related to the laser power This will add a systematic error to absolute power measurements This error may disappear from relative power measurements When you subtract two measurements m...

Страница 13: ...0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 10 FIG 1 2 TYPICAL GE TEMPERATURE DEPENDENCE VS THE WAVELENGTH FIG 1 3 TYPICAL SI AND SIUV TEMPERATURE DEPENDENCE VS THE W...

Страница 14: ...saturate the detector To know if the detector is saturated use procedure 4 3 1 Be careful when making the attenuator calibration in pulse mode The peak power must be in the linear region of the photo...

Страница 15: ...esired Wavelength The difference between the desired wavelength and the inferior wavelength 4 5 1 Example You have a PH100 Si and your laser is at 632 8 nm See your CERTIFICATE OF CALIBRATION for the...

Страница 16: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 13 FIG 1 4 PH100 SI TYPICAL SPECTRAL RESPONSE FIG 1 5 PH100 SIUV PH10B SI PE10B SI TYPICAL SPECTRAL RESPONSE...

Страница 17: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 14 FIG 1 6 PH20 GE PH5B GE PE5B GE TYPICAL SPECTRAL RESPONSE...

Страница 18: ...center when the detector reaches its end of life Breaking the calibration seal or opening the monitor will void the detector warranty The complete Detector contains 1 Detector with wires or DB 15 1 i...

Страница 19: ...Version 2 0 PH Series Instruction Manual Gentec Electro Optics Inc All rights reserved 16...

Отзывы: